Press "Read" to enter the corresponding parameter settings page, press "Store" /
"Mode" to adjust the parameters.
2.4.1 Standard buffer solution groups setting
Display “BUF”, press “Read” enter into standard buffer solution setting mode.
Note
:
Please refer to Appendix 5.1 for a total of 8 groups of buffer sets.
Default
:
1
(
NIST
)
2.4.2 Manual temp-compensation setting
Display “MTC”
,
press “Read” enter into manual temp-compensation setting mode.
Note
:
When the meter does not detect the temperature sensor, it will
automatically switch to manual temperature compensation mode, and display
MTC.
Default
:
25
℃。
2.4.3 Temperature unit setting
Display “TU”
,
press “Read” enter into temperature unit setting mode.
Default
:℃。
2.4.4 Prompt sound setting
Display ”BUZ”
,
press “Read” enter into prompt sound setting mode
Default:ON (with sound prompt).
2.4.5 Remove stored data
Display“MR CLr”
,
press “Read” enter into remove the stored data mode.
Display “Read” to confirm
,
display “MR nUL”
,
indicates clear success
,
Automatically exit this setting; press "ESC" to exit this setting.