BQ610-B Online Controller Operation manual
10
Trouble Shooting
Table 5-1 lists the symptoms, possible causes, and recommended solutions for
common problems encountered with the OIW sensor. If your symptom is not listed, or
if none of the solutions solves your problem, please contact us.
Table 5-1
ERROR
POSSIBLE CAUSE
SOLUTION
Communication
abnormal
Power supply or wiring
issues
Check whether the power supply and wiring are correct
according to the instruction
Interface or protocol
issues
1. Use our SmartPC upper computer software to check
whether the communication is normal
2. Check according to the supporting communication protocol
of the product
No change in
reading
Cleaning brush failure
1. Check whether the brush is entangled by foreign matter, if
so, please remove the foreign matter;
2. Turn on the power again and observe whether the brush
rotates. If it cannot rotate or rotates abnormally, please contact
customer service
Check whether the power supply meets the requirements, to
avoid the low power supply can not drive the brush rotation
Hardware or software
issues
Contact customer service
Measured value
is too high, too
Low or
instability
Sensor’s window is
dirty and worn
Clean sensor body, special light window table
Sensor’s brush is worn
Change brush
Calibration is required
Perform user calibration
Other errors
Contact customer service