BQ610-B Online Controller Operation manual
7
Installation
Part List
Item
Number
Note
OIW Sensor
1
Cable
1
10m
Brush
1
Assembly adapter
1
Sensor Installation
1) Wiring and power supply:
•
Do not use the sensor cable to pull the sensor! It is required to install sensor in a
secure and stable mounting bracket.
•
The female and male connector of sensor cable should be screwed tightly to avoid
moisture incursion
•
Important
: Make sure power supply voltage is correct before turning the power
on.
Sensor installation:
•
It is recommended to install the sensor vertically with
probe
facing down.
•
Considering the basic principle of sensor optics, the distance between the end
face of the sensor optical window and the bottom of the container/device should
be more than 10cm!
•
In consideration of the fluctuation of water level, install the sensor below water
level of 30cm to submerge, and try to install it in the position where there are no
bubbles in the water;
•
The sensor must be securely installed to avoid damage caused by water flow and
other things.