background image

BQ610-B Online Controller Operation manual

7

Installation

Part List

Item

Number

Note

OIW Sensor

1

Cable

1

10m

Brush

1

Assembly adapter

1

Sensor Installation

1) Wiring and power supply:

Do not use the sensor cable to pull the sensor! It is required to install sensor in a
secure and stable mounting bracket.

The female and male connector of sensor cable should be screwed tightly to avoid
moisture incursion

Important

: Make sure power supply voltage is correct before turning the power

on.

Sensor installation:

It is recommended to install the sensor vertically with

probe

facing down.

Considering the basic principle of sensor optics, the distance between the end
face of the sensor optical window and the bottom of the container/device should
be more than 10cm!

In consideration of the fluctuation of water level, install the sensor below water
level of 30cm to submerge, and try to install it in the position where there are no
bubbles in the water;

The sensor must be securely installed to avoid damage caused by water flow and
other things.

Summary of Contents for BQ-OIW

Page 1: ...BQ610 B Online Controller Operation manual 1 BQ OIW Online Oil in water Sensor User Manual SHANGHAI BOQU INSTRUMENT CO LTD...

Page 2: ...rview 4 Features 4 Introduction 4 Wire Definition 6 Technical Specifications 6 Installation 7 Part List 7 Sensor Installation 7 Calibration 8 Labware and materials required 8 Maintenance 9 Maintenance...

Page 3: ...or maintenance other than which mentioned in the manual Due to non compliance with the precautions specified in this manual any fault and loss caused shall not be covered by the warranty and the manuf...

Page 4: ...measurement data make the maintenance period longer and maintain excellent stability for long term online use Typical applications include crude and refined oil detection leakage control at offshore o...

Page 5: ...BQ610 B Online Controller Operation manual 5...

Page 6: ...nce wavelength 360nm Range 5ppm NDSA equiv to 1 1ppm BTEX Resolution 0 01ppb MDL 5ppb NDSA 1ppb BTEX Linearity R 0 999 Housing IP Rating IP68 Depth Rating 30m Max Temperature Range 0 50 Interface Supp...

Page 7: ...ke sure power supply voltage is correct before turning the power on Sensor installation It is recommended to install the sensor vertically with probe facing down Considering the basic principle of sen...

Page 8: ...in deionized water transfer it to volumetric flask constant volume to 1L and get 100ppm NDSA stock solution Different concentrations of standard solution preparation Dilute 100ppm NDSA stock solution...

Page 9: ...etergent can be added in tap water to clean 2 Window surface clean the outer surface of the sensor with tap water For some stubborn dirt traditional detergent and soft cloth can be used to clean it It...

Page 10: ...2 Check according to the supporting communication protocol of the product No change in reading Cleaning brush failure 1 Check whether the brush is entangled by foreign matter if so please remove the f...

Page 11: ...jeure natural disasters social unrest war published or unpublished terrorism civil war or any government forced Damage caused due to improper use negligence accident or caused by the improper applicat...