Operating Manual of Ammonia Nitrogen Sensor
8
Chapter 3 Installation
3.1Sensor connection
3.1.1 Specific steps for instrument connection
Remove the protective caps before powering the instrument, and then connect the cable core correctly
according to the core definition table.
Important Tip:
The power supply of this product is 12VDC. It is not allowed to connect directly to
220VAC or 24VDC power supply. Otherwise, the sensor will be over-pressed and damaged. If there is no
suitable power supply at the site, you can use the 220VAC to 12VDC power adapter that is equipped with
the instrument and then connect the instrument.
Core Definition Table:
Core No.
1
2
3
4
5
Sensor Wire
Brown
Black
Blue
White
Yellow
Signal
+12VDC
AGND
RS485 A
RS485 B
CHGND
Connection with software: The ammonia nitrogen sensor has standard MODBUS protocol and
RS-485 interface, which can be directly connected to the host computer software to realize real-time
online monitoring and data transmission of the instrument.