BOONTON ELECTRONICS a WIRELESS TELECOM GROUP COMPANY
21
e. Zero the reference powermeter as follows:
Pressing the
Zero/CAL
key.
Select the appropriate channel by pressing
<
or
>
.
Select
Zero Chan
and press
Enter
, and wait for the zero process to complete.
OR
Send “CAL1:INTERNAL:ZERO” to the
Meas4530
.
f. Configure the measurement and display settings on the
Meas4530
:
Select
Main Menu > Channel 1 > Parameters > Filter time
4 secs
Select
Main Menu > Channel 1 > Parameters > Frequency
0.05GHz
Select
Main Menu > Channel 1 > Display > Units
Log-dBm
Select
Main Menu > Channel 1 > Display > Display Source
Sensor1
OR
Send “SENSE1:FILTER:TIME 4; CALC1:UNITS DBM” to the
Meas4530
Send “SENSE1:CORR:FREQ 0.05; CALC1:MODE CW; CALC1:MATH CH1” to the
Meas4530
g. Set the 2520’s RF output to the first power step (-60dBm) with the RF output “
ON
”.
h.
[Set Ref]
to load the 2520’s current output as a reference level.
i.
[Select UUT]
and send “OUTPUT:INT:LEVEL –60; OUTPUT:INT:SIGNAL ON” to the
UUT
.
j.
Measure and record the error between the reference calibrator’s level and the output of the UUT in the
appropriate row of the table below.
k. Repeat steps (g) through (j) for each power level in the following table.
Level
Error
+
/- Limit
-60.0 dBm _______dBr
0.300
-59.9 dBm _______dBr
0.300
-50.0 dBm _______dBr
0.100
-49.9 dBm _______dBr
0.100
-40.0 dBm _______dBr
0.015
-39.9 dBm _______dBr
0.015
-30.0 dBm _______dBr
0.015
-29.9 dBm _______dBr
0.015
-20.0 dBm _______dBr
0.015
-19.9 dBm _______dBr
0.015
-10.0 dBm _______dBr
0.015
-9.9 dBm _______dBr
0.015
0.0 dBm _______dBr
0.015
0.1 dBm _______dBr
0.015
10.0 dBm _______dBr
0.015
10.1 dBm _______dBr
0.015
20.0 dBm _______dBr
0.015
Differences in the error between adjacent 0.1dB steps indicate the incremental error associated with
each 10dB range change. Errors in the fullscale and downscale D/A coefficients may show here.
l.
Remove the coaxial switch and adapters, and connect the sensor directly to the UUT’s calibrator.
m. Clear the reference and go back to measuring absolute power as follows:
Press the
Text
key twice. This displays a menu at the top of the display. Scroll left or right until
Ld/ClrR
is selected then press the
DN
arrow key.
Scans by ArtekMedia => 2009
Summary of Contents for 4530 Series
Page 25: ...BOONTON ELECTRONICS a WIRELESS TELECOM GROUP COMPANY 24 Scans by Artekmedia 2009 ...
Page 26: ...BOONTON ELECTRONICS a WIRELESS TELECOM GROUP COMPANY 25 Scans by ArtekMedia 2009 ...
Page 27: ...BOONTON ELECTRONICS a WIRELESS TELECOM GROUP COMPANY 26 Scans by Artekmedia 2009 ...
Page 29: ...BOONTON ELECTRONICS a WIRELESS TELECOM GROUP COMPANY 28 Scans by Artekmedia 2009 ...
Page 31: ...BOONTON ELECTRONICS a WIRELESS TELECOM GROUP COMPANY 30 Scans by Artekmedia 2009 ...
Page 32: ...BOONTON ELECTRONICS a WIRELESS TELECOM GROUP COMPANY 31 Scans by ArtekMedia 2009 ...