3 Safety instructions
bon E-20 SE
3 Safety instructions
Please adhere to the legal guidelines for accident preventions and take note of the following
safety instructions.
Set up and Installation:
7/23
The Refraction Unit must not be installed or used in damp rooms.
Please ensure the refraction unit is mounted on a level and stable surface.
The power must correlate to that shown on the Type label.
The maximum carrying capacity of the translation table (6) is 40 kg, 25 kg maximum can be placed in
position 1 of the translation table.
To Operate:
Do not expose the system to any temperature extremes. The recommended usage temperature lies
b 10° and + 40° C.
Avoid dripping water and water spray.
The maximum carrying capacity of the patient chair (5) is 170 kg. This must not be exceeded.
When the translation table (6) and the contact lens fitting table (7) are swung into position. There is a
scissor movement. This presents a risk of trapping!
The elevation column of the patient chair is not designed for continuous use. After continuous use for 60
seconds a cooling off period of 9 minutes must be adhered to.
Please note that neither the translation table (6) nor the contact lens fitting (7) should be allowed to hit
the patient` legs.
Do not exert any unnecessary pressure on the contact lens table (e.g. Leaning with the hand).
Ensure that the phoropter arm (2) does not hit the patient (e.g. as the chair moves upwards).
Longer use of the examination lamp (3) can lead to a heating of the lamp housing. There is a danger of
burning.
Other:
Do not pull the cable when removing the plug from the socket.
Position the cable so that there is no risk of walking on it or falling over it. Any other risk of damage to the
cable must be avoided (e.g. sharp edges or strong heat).
The plug must be removed from the socket before carrying out repairs, maintenance or commissioning.
Fig. 3-1: Refraction Unit bon E-20 SE