3  Safety instructions 

 

 

 

bon E-20 SE 

3 Safety instructions 

Please adhere to the legal guidelines for accident preventions and take note of the following 
safety instructions. 

 

Set up and Installation: 

 

 

 

7/23

 

 

 

ƒ

 

The Refraction Unit must not be installed or used in damp rooms. 

ƒ

 

Please ensure the refraction unit is mounted on a level and stable surface. 

ƒ

 

The power must correlate to that shown on the Type label. 

ƒ

 

The maximum carrying capacity of the translation table (6) is 40 kg, 25 kg maximum can be placed in 
position 1 of the translation table. 

 

 

To Operate: 

 

ƒ

 

Do not expose the system to any temperature extremes. The recommended usage temperature lies  

       b 10° and + 40° C. 

ƒ

 

Avoid dripping water and water spray. 

ƒ

 

The maximum carrying capacity of the patient chair (5) is 170 kg. This must not be exceeded. 

ƒ

 

When the translation table (6) and the contact lens fitting table (7) are swung into position. There is a 
scissor movement. This presents a risk of trapping! 

ƒ

 

The elevation column of the patient chair is not designed for continuous use. After continuous use for 60 
seconds a cooling off period of 9 minutes must be adhered to. 

ƒ

 

Please note that neither the translation table (6) nor the contact lens fitting (7) should be allowed to hit 
the patient` legs.  

ƒ

 

Do not exert any unnecessary pressure on the contact lens table (e.g. Leaning with the hand). 

ƒ

 

Ensure that the phoropter arm (2) does not hit the patient (e.g. as the chair moves upwards). 

ƒ

 

Longer use of the examination lamp (3) can lead to a heating of the lamp housing. There is a danger of 
burning. 

 
 

Other: 

 

ƒ

 

Do not pull the cable when removing the plug from the socket. 

ƒ

 

Position the cable so that there is no risk of walking on it or falling over it. Any other risk of damage to the 
cable must be avoided (e.g. sharp edges or strong heat). 

ƒ

 

The plug must be removed from the socket before carrying out repairs, maintenance or commissioning.

 

 

 

 

 

Fig. 3-1: Refraction Unit bon E-20 SE 

Summary of Contents for E-20 SE

Page 1: ...de Internet www bon de Sparkasse zu L beck Sort code 230 501 01 Acc No 1 014 885 Swift BIC HSHN DE H1 SPL IBAN DE 2305 0101 0001 0148 85 Commerzbank L beck Sort code 230 400 22 Acc No 0 107 755 Postb...

Page 2: ...DS 2 23...

Page 3: ...e 9 5 2 Commissioning 9 5 3 Description and function of the components 10 5 3 1 Translation table 10 5 3 2 Patient chair 11 5 3 3 Phoropter arm 11 5 3 4 Contact lens fitting table with trial lens tray...

Page 4: ...n the operating instructions Caution Please observe safety instructions with this symbol to prevent personal danger or damage to property Important Indicates particularly important information to main...

Page 5: ...product class 1 2 3 Liability The Refraction Unit is manufactured according to the recognized technical safety regulations and is tested to the strictest quality measures The company bon Optic accepts...

Page 6: ...onfiguration purchased Fig 2 1 bon E 20 SE with maximum parts and accessories 2 Pieces of Table connections for the table top devices only supplied when a table is supplied 1 Mains cable 1 5 m 1 Instr...

Page 7: ...table 7 are swung into position There is a scissor movement This presents a risk of trapping The elevation column of the patient chair is not designed for continuous use After continuous use for 60 s...

Page 8: ...dical devices which are to be connected to the electrical supply of the refraction unit must be proved to fulfill their DIN EN as well as IEC specifications All configurations must fulfill the require...

Page 9: ...nt chair 9 23 Fig 5 1 Switches on the control console 5 2 Commissioning 1 Please ensure that the refraction unit is connected to the power supply with the intended power cable 2 Press the On Off switc...

Page 10: ...8 Technical Data Power will be supplied automatically to the table top device on switching into working position 1 or 2 The translation table ensures a via a potentiometer the smooth introduction of...

Page 11: ...oropter back into the rest position 11 23 Position 1 Rest position Position 2 Phoropter arm angled Position 3 Phoropter arm in extamination position Fig 5 3 Phoropter Arm Positions The phoropter arm s...

Page 12: ...6 3 Repairs that can be carried out by the user The following repairs can be carried out by the user 6 3 1 Changing the bulb in the examination lamp Bulb used Halogen mirror lamp with reflector and sa...

Page 13: ...fects Component Defect Fuse Fuse Value Chair elevation Up and down movement of patient chair does not function F1 T 3 15 A Translation table Examination instruments The table cannot be locked the inst...

Page 14: ...ns apply in both instances The invoice with the purchase date applies The device was used according to the instructions and for the specified purpose Any repairs were carried out by bon personnel or p...

Page 15: ...tion unit bon E 20 SE View from above Other Dimensions Width Base plate 96 cm Depth Base plate 66 cm Height without projector standard components 165 cm Height Translation table 85 cm Seat Height At l...

Page 16: ...20 W Protection class I Device type B Protection type IP 21 Maximum Output Power Output 1 Output 2 88 W Conditions of Use Refraction Unit Ambient temperature 10 C to 40 C Relative air humidity 30 to 7...

Page 17: ...ge for tabletop unit 2 6 V AV 12 V AC 70 E_WTEIN Switching voltage 12 V AC 71 GND Switching voltage Ground 72 12 V DC Switching voltage 12 V DC 73 10VAC SB Operating voltage 10 V AC 74 Operating volta...

Page 18: ...8 Technical data bon E 20 SE Fig 8 3 Circuit diagram 18 23...

Page 19: ...8 Technical data bon E 20 SE Fig 8 4 Connecting diagram refraction unit bon E 20 SE 19 23...

Page 20: ...8 Technical data bon E 20 SE Fig 8 5 Connecting diagram control console Fig 8 6 Connecting diagram fuse back control console 20 23...

Page 21: ...8 Technical data bon E 20 SE Fig 8 7 Connecting diagram adjustable table 21 23...

Page 22: ...of it Instead it should be handed over to an applicable collection point for the recycling of electrical and electronic equipment If you have any further questions please contact bon optic In the cas...

Page 23: ...1 Richtlinie 93 42 EWG Anhang IX Regel 1 1 MDD 93 42 EEC annex IX rule 1 Wir erkl ren hiermit die bereinstimmung des vorgenannten Produkts mit der EU Richtlinie 93 42 EWG ber Medizinprodukte We decla...

Reviews: