background image

 
6  Care and maintenance 

 

 

 

bon E-20 SE 

6.3.2 Safety fuse 

 Only use fuses with the safety values listed in illustration 6-1 Type (T)! 
 
 

 

 

13/23

 

 

 

 

Fig. 6-1: Fuses behind the refraction unit 

 
1)  Bevore changing the fuse removes the plug from the mains socket. 
2)  Open the relevant fuse holder with a screw driver. 
3)  Remove the old fuse from the holder and place a new fuse of the same type in the holder. 
4)  Push the fuse holder back into the fuse port. 
5)  Close the fuse holder with a screw driver. 
 

6.3.3 Function defects 

Component 

Defect 

Fuse 

Fuse Value 

Chair elevation  

Up and down movement of patient chair does not function

F1 

T 3,15 A  

Translation table  
(Examination instruments) 

The table cannot be locked, the instruments do not work. 

Relay „clicks“ upon activating the table lock switch 

F2 

T 4,0 A  

Translation table 
(Examination instruments) 

The table cannot be locked into position instruments do 
not work. 

Relay „clicks“ without table lock being activated

 

F3 

T 6,3 A  

Examination Light 

Examination light and key pad illumination do not function 

F4 

T 3,15 A 

Main fuse power supply 

Refraction unit does not function 

F5 

T 5,0 A  

Main fuse power supply 

Refraction unit does not function 

F6 

T 5,0 A  

Table 1: Defects 

 
 
 
 
 
 

Summary of Contents for E-20 SE

Page 1: ...de Internet www bon de Sparkasse zu L beck Sort code 230 501 01 Acc No 1 014 885 Swift BIC HSHN DE H1 SPL IBAN DE 2305 0101 0001 0148 85 Commerzbank L beck Sort code 230 400 22 Acc No 0 107 755 Postb...

Page 2: ...DS 2 23...

Page 3: ...e 9 5 2 Commissioning 9 5 3 Description and function of the components 10 5 3 1 Translation table 10 5 3 2 Patient chair 11 5 3 3 Phoropter arm 11 5 3 4 Contact lens fitting table with trial lens tray...

Page 4: ...n the operating instructions Caution Please observe safety instructions with this symbol to prevent personal danger or damage to property Important Indicates particularly important information to main...

Page 5: ...product class 1 2 3 Liability The Refraction Unit is manufactured according to the recognized technical safety regulations and is tested to the strictest quality measures The company bon Optic accepts...

Page 6: ...onfiguration purchased Fig 2 1 bon E 20 SE with maximum parts and accessories 2 Pieces of Table connections for the table top devices only supplied when a table is supplied 1 Mains cable 1 5 m 1 Instr...

Page 7: ...table 7 are swung into position There is a scissor movement This presents a risk of trapping The elevation column of the patient chair is not designed for continuous use After continuous use for 60 s...

Page 8: ...dical devices which are to be connected to the electrical supply of the refraction unit must be proved to fulfill their DIN EN as well as IEC specifications All configurations must fulfill the require...

Page 9: ...nt chair 9 23 Fig 5 1 Switches on the control console 5 2 Commissioning 1 Please ensure that the refraction unit is connected to the power supply with the intended power cable 2 Press the On Off switc...

Page 10: ...8 Technical Data Power will be supplied automatically to the table top device on switching into working position 1 or 2 The translation table ensures a via a potentiometer the smooth introduction of...

Page 11: ...oropter back into the rest position 11 23 Position 1 Rest position Position 2 Phoropter arm angled Position 3 Phoropter arm in extamination position Fig 5 3 Phoropter Arm Positions The phoropter arm s...

Page 12: ...6 3 Repairs that can be carried out by the user The following repairs can be carried out by the user 6 3 1 Changing the bulb in the examination lamp Bulb used Halogen mirror lamp with reflector and sa...

Page 13: ...fects Component Defect Fuse Fuse Value Chair elevation Up and down movement of patient chair does not function F1 T 3 15 A Translation table Examination instruments The table cannot be locked the inst...

Page 14: ...ns apply in both instances The invoice with the purchase date applies The device was used according to the instructions and for the specified purpose Any repairs were carried out by bon personnel or p...

Page 15: ...tion unit bon E 20 SE View from above Other Dimensions Width Base plate 96 cm Depth Base plate 66 cm Height without projector standard components 165 cm Height Translation table 85 cm Seat Height At l...

Page 16: ...20 W Protection class I Device type B Protection type IP 21 Maximum Output Power Output 1 Output 2 88 W Conditions of Use Refraction Unit Ambient temperature 10 C to 40 C Relative air humidity 30 to 7...

Page 17: ...ge for tabletop unit 2 6 V AV 12 V AC 70 E_WTEIN Switching voltage 12 V AC 71 GND Switching voltage Ground 72 12 V DC Switching voltage 12 V DC 73 10VAC SB Operating voltage 10 V AC 74 Operating volta...

Page 18: ...8 Technical data bon E 20 SE Fig 8 3 Circuit diagram 18 23...

Page 19: ...8 Technical data bon E 20 SE Fig 8 4 Connecting diagram refraction unit bon E 20 SE 19 23...

Page 20: ...8 Technical data bon E 20 SE Fig 8 5 Connecting diagram control console Fig 8 6 Connecting diagram fuse back control console 20 23...

Page 21: ...8 Technical data bon E 20 SE Fig 8 7 Connecting diagram adjustable table 21 23...

Page 22: ...of it Instead it should be handed over to an applicable collection point for the recycling of electrical and electronic equipment If you have any further questions please contact bon optic In the cas...

Page 23: ...1 Richtlinie 93 42 EWG Anhang IX Regel 1 1 MDD 93 42 EEC annex IX rule 1 Wir erkl ren hiermit die bereinstimmung des vorgenannten Produkts mit der EU Richtlinie 93 42 EWG ber Medizinprodukte We decla...

Reviews: