Recommended Separation Distances Between Portable and Mobile RF Communications 
Equipment and the 3600 Handpieces @ 3Vrms 
 

3600 Handpieces are intended for use in an electromagnetic environment in which radiated 
RF  disturbances  are  controlled.  The  customer  or  the  user  of  3600  Handpieces  can  help 
prevent  electromagnetic  interference  by  maintaining  a  minimum  distance  between 
portable  and  mobile  RF  communications  equipment  (transmitters)  and  the  3600 
Handpieces  as  recommended  below,  according  to  the  maximum  output  power  of  the 
communications equipment. 

 

Separation Distance According to Frequency of Transmitter (meters) 

Rated  Maximum 
Output  Power  of 
Transmitter 
(Watts) 


150  kHz  to  80 
MHz 

  3.5    

 

 

 

v

1     

80 MHz to 800 MHz 

  3.5    

 

E   

 

      1   

800 MHz to 2.5 GHz 

   7         

 

 

      1   

0.01 

0.12 

0.12 

0.23 

0.1 

0.34 

0.34 

0.74 

1.7 

1.7 

2.3 

10 

3.7 

3.7 

7.4 

100 

11.7 

11.7 

23.3 

For  transmitters  rated  at  a  maximum  output  power  not  listed  above,  the  recommended 
separation distances 

in meters (m) can be estimated using the equation applicable to the 

frequency of the transmitter, where 

is the maximum output power rating of the transmitter 

in watts (W) according to the transmitter manufacturer. 

Note  1:  At  80  MHz  and  800  MHz,  the  separation  distance  for  the  higher  frequency  range 
applies. 

 

Note  2:  These  guidelines  may  not  apply  in  all  situations.  Electromagnetic  propagation  is 
affected by absorption and reflection from structures, objects, and people. 

 

 

11 Transport and storage conditions 
 

1.ambient temperature range:

 

-10

40

   

2.relative humidity range: ≤90%   
3.Kpa: 500hPa

1060hPa   

 

12 Operating conditions

 

normal working conditions:   
ambient temperature

35

   

relative humidity

35%

75%   

Kpa: 860hPa

1060hPa   

Supply voltage

AC 110V/220V

tolerance±10%

frequency

50Hz

tolerance±1Hz 

motor power

:(

DC

12 V   

There is no conductive dust, explosive gas, and corrosive gases 

Summary of Contents for BJWJZ-1

Page 1: ...1 MICROTYPE SURGICAL POWER TOOLS System 3600 Operating manual...

Page 2: ...3600 Page 7 6 4 Bur attachments Page 8 6 5 Craniotomy attachment BJ3605 Page 8 6 6 Sagittal and reciprocating attachments BJ3601 BJ3609 Page 9 6 7 Dedicated cranial perforator BJ3604 Page 9 7 Instruct...

Page 3: ...handle 5000 series Note to simplify angled and straight attacks are collectively called bur attachments 2 Product classification Based on the classification of the protection against electric shock th...

Page 4: ...mmergere il dispositivo in liquidi 10PB In relazione a folgorazioni elettriche incen dio pericoli meccanici solo in conformit a EN 60601 1 e ANSI AAMI ES60601 1 2005 e CAN CSA C22 2 N 60601 1 2008 Que...

Page 5: ...ly c The bur is a consumable product and its duration is generally 10 uses d The control unit has a short circuit protection device When a short circuit fault occurs the device automatically interrupt...

Page 6: ...drilling BJ3604 4 Connection port for BJ36012 pedal board 5 On screen button to choose between activation by footswitch or by screen 6 Direction of rotation 7 Increase and decrease speed indication o...

Page 7: ...Foot pedal BJ36012 Cable 3 meters Water resistant IPX6 6 3 Micro handpiece BJ3600 Input voltage DC 12V Output power 100W Noise 65dB Image Micro handpiece BJ3600 1 Handpiece connector 2 Motor control...

Page 8: ...8 6 4 Bur attachments Speed 1000rpm 100000rpm Error 8 Torque 0 10 Nm Connection type standard round shank 2 35mm 6 5 Craniotomy attachment BJ3605 Speed 1000rpm 100000rpm Error 8 Torque 0 10 Nm...

Page 9: ...peed 1000rpm 15000rpm Error 8 Connection type for sagittal saw blades code BJ3501 Connection type for reciprocating saw blades code WF 6 7 Dedicated cranial perforator BJ3604 Speed 100rpm 1100rpm Torq...

Page 10: ...vation button again the motor will run in the current speed and direction 5 When the instrument is not in use please disconnect the power cord Note After following the instructions now described pleas...

Page 11: ...allow moisture to remain inside the cannulated parts in the selectors or in the electrical connection areas Do not rinse the handpiece from the front or use pressurized water to avoid damaging the sys...

Page 12: ...to obtain optimum performance Steam sterilization Wrapping method Cycle Minimum temperatureMaximum Temperature Minimum ExposureMinimum dry time Wrapped Pre vacuum 129 C 134 C 4 minutes 8 minutes Note...

Page 13: ...equipment contains no repairable parts inside so do not attempt to open the enclosure the foot switch the control box and other components If other problems persist contact the after sales service fo...

Page 14: ...unctions therefore RF emissions are very low and are not likely to cause any interference in nearby electronic equipment RF Emissions CISPR 11 Class A The 3600 Handpieces are intended for use by healt...

Page 15: ...vered with synthetic material the relative humidity should be at least 30 Electrical fast 2 kV for power supply lines 2 kV for power supply lines Mains power quality should transients bursts be that o...

Page 16: ...n 70 Ut 30 dip in Ut for 25 cycles 70 Ut 30 dip in Ut for 25 cycles during power mains interruptions it is recommended that the 3600 5 Ut 95 dip in Ut for 5 seconds 5 Ut 95 dip in Ut for 5 seconds Han...

Page 17: ...in each frequency range b Interference may occur near the equipment marked with the following symbol NOTE 1 At 80 MHz and 800 MHz the higher frequency range applies NOTE 2 These guidelines may not app...

Page 18: ...7 2 3 10 3 7 3 7 7 4 100 11 7 11 7 23 3 For transmitters rated at a maximum output power not listed above the recommended separation distances d in meters m can be estimated using the equation applica...

Page 19: ...ou need technical information such as circuit diagram list of components illustrations correction details etc you can request it at the Technical Assistance Center Shanghai Bojin Electric Instrument D...

Reviews: