16
Voltage dips, short
interruptions and
voltage variations
on power supply
<5%
Ut
(>95% dip in
Ut
)
for 0.5 cycle
<5%
Ut
(>95% dip in
Ut
)
for 0.5 cycle
Mains power quality
should be that of a
typical commercial or
hospital
input lines
IEC 61000-4-11
40%
Ut
(60% dip in
Ut
)
for 5 cycles
40%
Ut
(60% dip in
Ut
)
for 5 cycles
environment. If the
user of the 3600
Handpieces requires
continued operation
70%
Ut
(30% dip in
Ut
)
for 25 cycles
70%
Ut
(30% dip in
Ut
)
for 25 cycles
during power mains
interruptions, it is
recommended that
the 3600
<5%
Ut
(>95% dip in
Ut
)
for 5 seconds
<5%
Ut
(>95% dip in
Ut
)
for 5 seconds
Handpieces
be
powered from an
uninterruptable
power supply or
battery.
NOTE: Ut is the a.c. mains voltage prior to application of the test level.
Portable and mobile RF communications equipment should be no closer to any part of the
3600 Handpieces, including cables, than the recommended separation distance calculated
from the equation applicable to the frequency of the transmitter.