6.2 Turn the unit on and simultaneously press the Up and
Down arrow keys and the Program key to enter
calibration mode.
6.3 Allow the unit 60 minutes to achieve calibration
setpoint of 40.0°C. Once the 60 minutes have elapsed,
the unit will beep once and flash the 40.0°C text on the
display. Use the Up and Down arrow keys to match the
displayed value to the temperature of the calibrated
probe. Press Set key to confirm temperature reading.
6.4 Allow the unit 60 minutes to achieve new calibration
setpoint of 60.0°C. Once the 60 minutes have elapsed,
the unit will beep once and flash the 60.0°C text on the
display. Use the Up and Down arrow keys to match the
displayed value to the temperature of the calibrated
probe. Press Set key to confirm temperature reading.
6.5 The unit will return to the main screen and can be
turned off at this point. Allow the unit to cool before
touching any internal surfaces. Once this is done, the
probe setup can be removed and normal operation of
the slide hybridization oven can be resumed.