5.2.4.3

 

Press Set key again to toggle between 
adjusting set temperature and time 
remaining. 

 

5.2.4.4

 

Once all program steps are entered, pressing 
Start/Stop/Reset will begin the program. 
Pressing Start/Stop/Reset again will pause the 
program. Pressing and holding 
Start/Stop/Reset will reset the program to S1. 

 

5.2.4.5

 

Temperatures can be adjusted from ambient 
or 18°C (whichever is greater) to 75°C. 

 

5.2.4.5.1

 

S2 and S3 can be turned off in the 
program by adjusting the 
temperature value down to the 
value below 18.0°C, which will 

display the word “OFF”.

 

 

5.2.4.6

 

Times can be adjusted from 000:01 to 999:59 
(mmm:ss). 

 

5.2.4.6.1

 

S2 and S3 can be set to hold 
indefinitely in the program by 
adjusting the time value down to the 
value below 000:01, which will 

display the word “HOLD”.

 

 

5.2.4.7

 

Upon completion of all steps, unit will beep 
and flash time as 000:00. Temperature will 
remain at last set temp. Holding 
Start/Stop/Reset resets the cycle and shows S1 
with its time and temperature. 

 

6.0  Calibration 
 

6.1    Turn the unit off and allow the interior surfaces to reach 

ambient temperature. Place a certified reference 
thermometer in the center of the tray.  Ensure the 
thermometer is not touching any parts of the tray. 
Boekel calibration kit part number 240550 can be used 
to assist in the setup.   

 

Summary of Contents for RapidFISH 240200-2

Page 1: ...N2400362 Rev 1 04 16 RapidFISH Models 240200 and 240200 2 Operating Instructions...

Page 2: ...4 5 0 Operation 5 5 1 Turning the Unit On 5 5 2 Modes of Operation 6 6 0 Calibration 7 7 0 Technical Specifications 9 8 0 Additional Accessories Available for Purchase 10 9 0 Fault Diagnosis 10 10 0...

Page 3: ...hat it is difficult to operate the disconnecting device Use only as specified by the operating instructions or the intrinsic protection may be impaired After transport or storage in humid conditions d...

Page 4: ...ir replacement for shipping damages must be coordinated through the carrier Complete and return the Warranty Registration Card or you may also complete this Warranty Card online at www boekelsci com T...

Page 5: ...e or decrease the selected variable Press and hold for rapid change E LCD Display Shows set temperature program number actual temperature time remaining and heating indicator F When the unit is in ful...

Page 6: ...time and temperature values to be run sequentially 5 2 2 The Program key cycles through the different programs S1 S2 S3 HT S1 and so forth 5 2 3 In HT mode the RapidFISH will hold temperature at the d...

Page 7: ...ay the word OFF 5 2 4 6 Times can be adjusted from 000 01 to 999 59 mmm ss 5 2 4 6 1 S2 and S3 can be set to hold indefinitely in the program by adjusting the time value down to the value below 000 01...

Page 8: ...key to confirm temperature reading 6 4 Allow the unit 60 minutes to achieve new calibration setpoint of 60 0 C Once the 60 minutes have elapsed the unit will beep once and flash the 60 0 C text on the...

Page 9: ...with IEC 664 For operation at altitudes of up to 6500 feet 2000 meters Temperature Range Ambient 5 0 C to 75 C Stability 0 2 C Accuracy 0 5 C Temperature Display Resolution 0 1 C Supply Ratings Model...

Page 10: ...d a Actual temperature is higher than Set temperature b Temperature control circuit fault c Circulation Fan failure a Check set temperature b Have unit checked by a competent service person c Have uni...

Page 11: ...t if hazardous material is spilled onto or into the equipment no decontamination or cleaning agents are used which could cause a hazard as a result of a reaction with parts of the equipment or with ma...

Page 12: ...w to adjust the contrast up or down respectively Care should be taken to avoid adjusting contrast so far in one direction that the text on the display becomes unreadable 12 0 Warranty When used in lab...

Page 13: ...l Scientific prior to receiving the RMA number Please consult the manufacturer or his agent if there is any doubt about the compatibility of decontamination or cleaning agents Please be sure to mark t...

Reviews: