The nosepiece is not in the located 
position 

Turn the nosepiece in the required 

position 

The interpupillar distance is not 

correct 

Adjust the interpupillar distance 

correctly 

The diopter is not right 

Adjust the diopter according your 

sight 

 

6

 The eyes are 

uncomfortable, the 

left and right fields of 

view is not coincided . 

 

Can’t adapt to binocular 

observation   

When  look  into  the  objective,  do  not 
stare  at  the  specimen  but  at  the 
whole field of view, or move the eyes 
away  to  see  other  things,  then  back 
into the objective   

 

                                             

 

 

 

                                          BM-800 

PROBLEM 

REASON FOR PROBLEM 

SOLUTION 

II

Mechanical Part: 

1

 The coarse focus 

knob is hard to run 

The tension adjustment collar is too 

tight 

Loose properly 

2

 The image can’t 

stay on the focal 

plane in the 

process of the 

observation 

 

The tension adjustment collar is too 

loose 

Tighten properly 

III

Electric Part: 

No power supply 

Check the power cord, and connect 

them exactly 

the installation of the bulb is wrong 

Install the bulb correctly 

1

 The lamp can’t 

light 

The bulb burn out 

Change a new bulb 

The  bulb  burn  out  in  a 
high frequency 

Not use the specified lamp 

Use the required lamp   

Not use a appointed lamp 

use a appointed lamp 

2

 The height of the 

brightness is not 

enough   

The brightness adjustment knob is 

used wrong 

Adjust the brightness adjustment 

knob in a correct way 

The bulb is going to spoil 

Change the bulb 

3

 The light glimpse 

The power cord have a poor 

contact 

Check the power cord, and connect 

them exactly 

 

Summary of Contents for BM-800

Page 1: ...This manual is written for lab biological microscope BM 800 For safety exerting best performance of the instrument and making you familiar with the instrument entirely we strongly recommended that yo...

Page 2: ...Contents BM 800 Use Notices 1 Name of Components 1 2 Installation 2 2 1 Installation Diagram 2 2 2 Installation Steps 3 3 Adjustment 7 4 Operation 9 5 Technical Specifications 13 6 Trouble Shooting 14...

Page 3: ...tage or it will bring a serious damage to the instrument 9 Use the factory supplied power cord please II Maintenance 1 All the lenses have been well checked and adjusted It is forbidden to disassemble...

Page 4: ...1 Name of Components BM 800 Nosepiece Trinoclular Viewing Unit Objective Mechanical Stage Lamp House Video Port optional Eyepiece Ocular Microscope Body...

Page 5: ...nstallation sequence of the components The number in the figure show the installation steps Before installing be sure every components is clean do not score any parts or glass surface Keep well with t...

Page 6: ...Microscope Body Lamp House Mechanical Stage Support Device 10 Wide Field Eyepiece Infinite Plan Objective Trinocular Viewing Unit Video Port optional...

Page 7: ...2 2 2 Installation Steps BM 800 Guide Board 3 1 Locking Block and Bolt...

Page 8: ...l stage support device figure 2 place it from the top of the guide board figure 1 let the device figure2 falling free until it reach the limit position Use the hexagon wrench screw down the locking bl...

Page 9: ...amp holder 2 Replacing Lamp when using or soon after When using or soon after it is turned off the lamp the lamp house and nearby parts will be very hot and will cause serious burns Please turn the ma...

Page 10: ...epiece from the left or the right side figure 9 then push the nosepiece clockwise then place other objectives by the sequence of low to high magnification figure 10 Installing objective this way will...

Page 11: ...into the eyepiece tube until they are against each other The result is showing in the figure 14 2 2 7Installing the Video Port optional Insert the video port figure 12 into the trinocular unit figure...

Page 12: ...3 Adjustment BM 800 Diopter Ring Left Fine Focus Knob Left Coarse Focus Knob Clocking Set Condenser Focus Knob Video Port optional...

Page 13: ...o Port optional Interpupillar Distance Indicator Right Fine Focus Knob Tension Adjustment Collar Right Coarse Focus Knob Brightness Adjustment Knob Main Switch Field Diaphragm Swing out Condenser with...

Page 14: ...irection the voltage decline and the brightness weaken Using the lamp in a low voltage condition will prolong the use life 4 3 Adjust the Tension Adjustment Collar figure 17 The tightness of the tensi...

Page 15: ...ng the objective touch the specimen 4 5 Adjusting the Interpupillar Distance Figure 19 The interpupillar distance range 48mm 75mm When observing with two eyes hold on the left and right prism holder t...

Page 16: ...estroying the specimen 2 When using the video set Pull out the light path selector lever see figure25 observe with both eyes when the image is sharp you can see the pictures directly on the video scre...

Page 17: ...ture Diaphragm The aperture diaphragm is designed for the adjustment of the numerical aperture not for the brightness Generally reducing the diaphragm opening to 70 80 of the N A value of the respecti...

Page 18: ...he observed image of the field of view is sharp then open the field diaphragm let the image full of the field of view reduce the mixed light improving the quality of the image 4 10 Switching the Light...

Page 19: ...n lamp 6V30W Condenser Swing out condenser NA0 9 0 25 Configuration Table Viewing Head Compensation Free Trinocular Head Eyepiece Extra Wide Field Eyepiece Infinite plan objective 4 10 40 100 Objectiv...

Page 20: ...lens There are stains on the specimen Change the specimen 3 Find dust and stain in the field of view There are stains on the eyepiece Clean the eyepiece The objective damage Mend and correct the objec...

Page 21: ...he tension adjustment collar is too tight Loose properly 2 The image can t stay on the focal plane in the process of the observation The tension adjustment collar is too loose Tighten properly III Ele...

Reviews: