12
Purpose
The last stable value remains displayed for 5 seconds after removing the sample
from the pan.
Example:
Determine weight of oversized sample
Step
Key (or instruction)
Display
1. Select application program
(ZERO)
≥
2 sec
2. Select Display Lock (“
HLdon
”)
(F) repeatedly
3. Confirm application
(ENTER)
Symbol “ ” on the display:
application is active
4. If necessary: zero the balance/scale
(ZERO)
5. Place sample on balance/scale
6. Start application program
(F)
Symbol “
” flashes on the display:
when stability conditions are met
7. Unload the balance/scale:
the weight value remains displayed
for a further 5 seconds; or
8. Zero the balance/scale
(ZERO)
9. End the Display Lock application
(ENTER)
≥
2 sec
10. Reactivate Display Lock (if no other
(F)
application program has been selected)
11. Repeat procedure starting from Step 5.
Display Lock Mode