background image

11

Weighing in Percent

Purpose

This application program allows you to obtain weight readouts in percent which are
in proportion to a reference weight.

Example:

Determine an unknown percentage; store the weight on the

balance/scale as the reference percentage (100%)

Step

Key (or instruction)

Display

1. Select application program 

(ZERO) 

2 sec

2. Select Weighing in Percent 

(F) repeatedly

3. Confirm application

(ENTER)

Symbol “ “ on the display: 
application is active

4. Place empty container on the 

balance/scale

5. Zero the balance/scale

(ZERO)

6. Place the reference weight 

for 100% on the balance/scale

7. If desired, change the number 

of decimal places displayed: 100.0%, 
100.00% or 100% (factory setting)

(F) repeatedly

8. Confirm selected no. of decimal places

(ENTER)

9. Place unknown weight on the 

balance/scale

10. Toggle display between weight 

and percentage

(F) repeatedly

11. Unload the balance/scale

12. Start new Weighing in Percent

application or exit application

(ENTER) 

2 sec

13. Reactivate Weighing in Percent 

(if no other application program 
has been selected)

(F)

14. Repeat procedure starting from Step 5

.

Summary of Contents for BEB43

Page 1: ...BEB Series Serie BEB Electronic Precision Scales Balances Balanzas electrónicas de precisión Operating Instructions Manual de instrucciones 98648 013 84 ...

Page 2: ...2 English page 3 In cases involving questions of interpretation the German language version shall prevail Español página 20 En caso de interpretación la versión en lengua alemana será determinante ...

Page 3: ...iption of the Keys 8 Application Programs 9 Toggling between Weigh Units 9 Counting 10 Weighing in Percent 11 Display Lock Mode 12 Total Formulation 13 Specific Gravity 14 Calibration Span Adjustment 15 Advanced Features Setup Menu 16 Error Codes 17 Overview 18 Specifications 18 C Marking 19 3 Contents ...

Page 4: ...t with liquid Exposure to excessive electro magnetic interference can cause the readout value to change Once the disturbance has ceased the instrument can be used again in accordance with its intended purpose Installation It is recommended to connect only BOECO accessories and options as these are optimally designed for use with your balance scale Do not open the balance scale housing as this will...

Page 5: ...draft ring Level indicator and adjustable feet Installation Choose a location that is not subject to the following negative influences Heat heater or direct sunlight Drafts from open windows and doors Extreme vibrations during weighing Excessive moisture Setting Up the Balance Scale Place the components on the balance scale in the following order Reversible weighing pan Round glass draft ring on m...

Page 6: ...tall a shield for protection against drafts if necessary Installing the Battery not for model BEB43 Batteries are not included with the equipment supplied Use only commercially available 9 V batteries If you use a rechargeable battery recharge it with an external battery charger Lay the balance scale on its side Open the battery compartment remove the compartment cover Install the battery in the c...

Page 7: ...ce scale does not rock In most cases this will require several adjustment steps Removing Weights for Calibration Span Adjustment not for model BEB61 Grasp the tab to turn and pull out the weight compartment Always wear gloves or use tweezers when handling weights Follow instructions on page 15 for calibration span adjustment Warmup Time for Model BEB43 To ensure accurate results the scale must war...

Page 8: ...r another 2 seconds the bal ance scale shuts off automatically unless a key is pressed Description of the Keys ON OFF On off key switches the bal ance scale on and off or switches it to the standby mode Battery operation on turns backlight on off ZERO Zeros the balance scale press and hold 2 seconds enters application menu CAL Starts calibration adjustment F Starts an application pro gram scrollin...

Page 9: ...weigh unit F Menu code Unit Conversion factor Display 1 user Grams 1 00000000000 o 2 grams factory setting Grams 1 00000000000 g 4 carat Carats 5 00000000000 o 5 Pound Pounds 0 00220462260 lb 6 ounce Ounces 0 03527396200 oz 7 tryo Troy ounces 0 03215074700 ozt 8 tl hon Hong Kong taels 0 02671725000 tlk 9 tl sin Singapore taels 0 02645544638 tl 10 tl ta Taiwanese taels 0 02666666000 tl 11 grai Grai...

Page 10: ...e balance scale 5 Zero the balance scale ZERO 6 Place sample reference quantity 20 on the balance scale 7 Select sample reference quantity in increments of 1 1 2 3 99 or F repeatedly briefly or in increments of 10 10 20 30 100 F 2 sec 8 Confirm selected sample reference quantity ENTER 9 Place uncounted parts on balance scale 10 Toggle display between mean piece weight total weight and quantity F r...

Page 11: ...n is active 4 Place empty container on the balance scale 5 Zero the balance scale ZERO 6 Place the reference weight for 100 on the balance scale 7 If desired change the number of decimal places displayed 100 0 100 00 or 100 factory setting F repeatedly 8 Confirm selected no of decimal places ENTER 9 Place unknown weight on the balance scale 10 Toggle display between weight and percentage F repeate...

Page 12: ...e display application is active 4 If necessary zero the balance scale ZERO 5 Place sample on balance scale 6 Start application program F Symbol flashes on the display when stability conditions are met 7 Unload the balance scale the weight value remains displayed for a further 5 seconds or 8 Zero the balance scale ZERO 9 End the Display Lock application ENTER 2 sec 10 Reactivate Display Lock if no ...

Page 13: ...flashes If interface is connected sample weight is outputted automatically 7 Remove sample or press ZERO 8 Place the next sample on the balance scale in this example 575 g 9 Store value in memory Total weight is ENTER displayed along with the flashing icon Note symbol remains on indicating stored value in memory until cleared if interface is connected sample weight is outputted automatically 10 To...

Page 14: ...display application is active 4 If necessary zero the balance scale ZERO 5 Start application program F 6 Confirm the display airval ENTER 7 Determine weight of the sample in air place sample on the balance scale 8 Store weight in air value ENTER 9 Remove sample from balance scale 10 Determine weight of sample in liquid connect wire and set up beaker 11 Confirm the display Water ENTER 12 Place samp...

Page 15: ...tch on the balance scale ON OFF 2 Zero the balance scale ZERO 3 Start calibration CAL The preset calibration weight is displayed without the weigh unit in this example 5000 g 4 To select a different calibration weight value F repeatedly 5 Start calibration span adjustment ENTER After the zero point is stored the required calibration weight flashes on the display 6 Place the required calibration we...

Page 16: ...ree numbers of a code hold to confirm setting ZERO Moves to the previous number of a codes hold to exit set up Parameter Settings Overview ο Factory setting Setup 1 Weighing 1 1 Adapt filter 1 1 1 Very stable conditions menu 1 1 2 ο Stable conditions 1 1 3 Unstable conditions 1 1 4 Very unstable conditions 1 2 Application filter 1 2 1 ο Final readout 1 2 2 Filling 1 3 Stability range 1 3 1 1 4 dig...

Page 17: ...crease the weight no sample on the balance scale on the balance scale with application in use PrtErr Data interface for Contact the BOECO printing is blocked customer service center Bal Err Balance scale loaded or defective Unload balance scale before when power was turned on switching on or contact BOECO customer service Sep Err Balance scale defective Contact BOECO customer service Max weighing ...

Page 18: ...115 V 15 to 20 48 60 Hz Power consumption average W 1 0 75 0 75 Hours of operation w 9V battery Alkaline approx h 11 14 Rechargeable fully ch NiMH avg h 2 5 4 Model BEB61 Weighing capacity g 5000 Readability g 0 1 Tare range subtractive g 5000 Linearity g 0 2 Operating temperature range 10 C to 30 C 273 K to 303 K 50 F to 86 F Response time average s 2 Adaptation to ambient conditions By selection...

Page 19: ...orrect these modifications On request BOECO will provide information on the minimum operating specifications in accordance with the Standards listed above for defined immunity to interference 73 23 EEC Electrical equipment designed for use within certain voltage limits Applicable European Standards EN 60950 Safety of information technology equipment including electrical business equipment EN 61010...

Page 20: ...sica Pesar 25 Descripción de teclas 25 Programas de aplicación 26 Conmutación de unidades 26 Conteo 27 Pesada en porcentaje 28 Lectura estática 29 Totalización 30 Densidad Peso específico 31 Calibración 32 Ajustes previos menú Setup 33 Mensajes de error 34 Sinopsis 35 Especificaciones técnicas 35 Marca C 36 ...

Page 21: ...ro tipo de pila batería puede causar explosión La balanza puede ser conectada solamente después de desconec tar el alimentador o desactivar la batería es decir exenta de tensión Proteger el alimentador contra la humedad Perturbaciones electromagnéticas extremas pueden influir en la lectura de los valores de peso Después de eliminarse las pertur baciones el producto funcionará nuevamente en la form...

Page 22: ...l de burbuja y patas de regula ción Instalación En la instalación evitar lugares con las siguientes influencias negativas calor calefacción rayos solares corrientes de aire directas debido a puertas y ventanas abiertas vibraciones durante el pesaje humedad extrema Instalar la balanza Montar los siguientes componentes en la balanza plato de carga redondo con la cara inferior o superior Protectores ...

Page 23: ... de aire Instalar pila batería excepto en los modelos BEB43 Pila o batería no forman parte del suministro de la balanza Usar solamente pila o batería monoblock de 9 V adquirible en el comercio Cargar batería solamente con un aparato recargador externo Voltear la balanza Abrir el compartimento de pila levantar la cubierta del comparti mento Colocar la pila o batería de 9 V Observar la polaridad Cer...

Page 24: ...e del nivel esté en el centro del círculo Por lo general se necesitan algunos pasos más para nivelar Sacar las pesas Excepto en el modelo BEB61 Girar la tapa del compartimento y jalar Proceso de calibración ver pág 32 Calentamiento previo para modelo BEB43 Para alcanzar resultados exactos el aparato necesita un tiempo de calentamiento previo de 1 hora Hasta entonces alcanza la temperatura necesari...

Page 25: ...nza se apaga automáticamente Descripción de teclas ON OFF Encender apagar el aparato se enciende se apaga o bien pasa al modo Standby Modo batería encender retroi luminación apagar ZERO Tarar balanza mantener pulsa da la tecla por 2 segundos ini ciar el menú de aplicaciones CAL Iniciar calibración F Iniciar programas de aplicación En el menú de aplicaciones y en calibración seleccionar parámetros ...

Page 26: ...stra 7 conmutar unidad peso F Num de menú Unidad Conversión Lectura 1 user Gramos 1 00000000000 o 2 grams ajuste fábrica Gramos 1 00000000000 g 4 carat Quilates 5 00000000000 o 5 Pound Libras 0 00220462260 lb 6 ounce Onzas 0 03527396200 oz 7 tryo Onzas Troy 0 03215074700 ozt 8 tl hon Taeles Hongkong 0 02671725000 tl 9 tl sin Taeles Singapur 0 02645544638 tl 10 tl ta Taeles Taiwan 0 02666666000 tl ...

Page 27: ...va 4 colocar recipiente en la balanza 5 tarar ZERO 6 colocar cantidad referencial 20 piezas 7 seleccionar cant referencia repetidam en escalones de uno 1 2 3 99 F pulsión breve en escalones de diez 10 20 30 100 o bien mantenerla F pulsada 8 confirmar cant ref seleccionada ENTER 9 poner cant desconocida de piezas 10 conmutar lectura entre promedio peso de pieza peso cant piezas repetidam F 11 desca...

Page 28: ...ajuste ENTER símbolo en la lectura aplicación está activa 4 colocar recipiente en la balanza 5 tarar ZERO 6 colocar peso referencial para el 100 7 en caso dado modificar decimal para lectura porcentaje 100 0 100 00 o bien 100 ajuste fábrica repetidam F 8 confirmar posic decimal seleccionada ENTER 9 poner peso desconocido 10 conmutar lectura entre peso y cantidad porcentaje repetidam F 11 descargar...

Page 29: ... ENTER símbolo en la lectura aplicación está activa 4 en caso dado tarar ZERO 5 poner muestra 6 iniciar aplicación F símbolo parpadea en la lectura valor peso estable fijo en la lectura 7 descargar balanza valor peso estable permanece aún 5 seg en la lectura 8 o bien tarar balanza y borrar valor estable indicado ZERO 9 finalizar aplicación Lectura estática Enter pulsión larga 10 reactivar Lectura ...

Page 30: ... aquí p ej 380 g 6 guardar valor en memoria total ENTER fijo en la lectura símbolo parpadea 7 quitar muestra 8 poner siguiente muestra en la balanza aquí p ej 575 g 9 guardar valor en memoria ENTER total fijo en la lectura símbolo parpadea 10 indicar componente actual por 3 seg F con impresora conectada impresión simultánea 11 borrar memoria totalización F pulsión larga con impresora conectada imp...

Page 31: ...ra aplicación está activa 4 en caso dado tarar balanza ZERO 5 iniciar aplicación F 6 confirmar lectura airval ENTER 7 determinar peso de muestra en el aire poner muestra en la balanza 8 memorizar valor peso en aire ENTER 9 Quitar muestra 10 determinar valor peso en líquido poner muestra en el canastillo 11 poner muestra en el líquido 12 confirmar lectura water ENTER 13 memorizar valor peso en líqu...

Page 32: ...seg Calibrar la balanza en el lugar cada vez que se cambie de lugar de instalación Ejemplo calibrar balanza aquí modelo BEB61 Paso Pulsar tecla Lectura 1 en caso dado encender bal ON OFF 2 tarar balanza ZERO 3 iniciar proceso calibración se indica CAL pesa calibración predefinida sin unidad aquí p ej 5000 g 4 en caso dado seleccionar valor pesa cal repetidam F 5 confirmar valor pesa cal e iniciar ...

Page 33: ... ambiente intranquilo 1 1 4 ambiente muy intranquilo 1 2 Filtro aplicación 1 2 1 ο pesar 1 2 2 dosificar 1 3 Rango estabilidad 1 3 1 1 4 de dígito 1 3 2 1 2 dígito 1 3 3 1 dígito 1 3 4 ο 2 dígitos 1 3 5 4 dígitos 1 5 Calibrar linealizar 1 5 1 ο calibrar función tecla CAL 1 5 2 linealizar sólo para el servicio 1 5 3 tecla bloqueada 1 6 Auto cero 1 6 1 ο activado 1 6 2 desactivado 1 7 1 Unidad peso ...

Page 34: ... no es representable calErr No se observó condición Quite carga de la balanza para calibrar p ej Tare con tecla ZERO puesta a cero Calibre recién con lectura cero plato aún con carga appErr Peso muy liviano o bien plato Aumente el peso sin carga en programas aplicación PrtErr Interfaz de datos para Comunique al servicio técnico imprimir bloqueada BOECO Bal Err Aparato con carga al encender Antes d...

Page 35: ...V o bien 115 V 15 20 48 60 Hz Consumo eléctrico típico W 1 0 75 0 75 Tiempo funcionam con pila monobloque 9 V alcalina aprox h 11 14 batería carga compl NiMH típico h 2 5 4 Modelo BEB61 Rango de pesada g 5000 Legibilidad g 00 1 Rango tara substract g 5000 Desv linealidad g 0 2 Rango temp de uso 10 C 30 C 273 K 303 K 50 F 86 F Tiempo estabilización típico s 2 Adaptación a las condic 4 escalas optim...

Page 36: ...r controladas por él mismo y en caso dado corregidas BOECO pone a disposición según consultas informaciones sobre la calidad de funcionamiento según las normas sobre inmunidad definida más arriba mencionadas 73 23 CEE Material eléctrico a utilizarse con determinados límites de tensión Normas Europeas aplicables EN 60950 Seguridad de los equipos de tratamiento de la información incluyendo los equip...

Page 37: ......

Page 38: ......

Page 39: ......

Page 40: ...Boeckel Co Roedingsmarkt 33 20459 Hamburg Germany t 49 0 40 3256270 f 49 0 40 32562741 E mail Boeckel Co t online de 98648 013 84 Publication No WXX6072 al05101 W1A000 BOECO BEB ...

Reviews: