5
4). Skin might get hurt if it absorbs too much light energy;
5). Put the handpiece on the handpiece stand if you don’t use it;
6). Keep the handpiece crystal clean and do not make cooling gel leak into handpiece;
7). Use cooling gel to protect the skin while treating.
2.4. Fire Prevention
1). The clinic room shall not be easy to get fire;
2). If you cleaned the equipment with alcohol, you shall not turn on the equipment until
it is totally dry;
3). If you don’t use this equipment for a long time, cut off the power supply.
2.3. Electrical Safety
2.5. Safety Device
Air Switch
:
If the electricity is abnormal, this electricity breaker would automatically
cut off electricity
;
Emergency Button
:
Press it and power is cut off. Turn it clockwise to connect power;
Key switch
:
It is used to turn on and turn off power;
1). The power supply shall be AC230V±10%,50Hz±1Hz/AC110V±10%,60Hz±1Hz;
2). Before maintenance, turn off equipment and wait for 10 minutes;
3). No one is allowed to disassemble any parts of the equipment except the technical
personnel appointed by our company;
4). If you find the handpiece leaking water, don’t turn on the equipment. If it has been
turned on, please turn it off;
5). The power line of the this equipment could conduct unexpected electricity to earth;
6). People in the clinical room shall know how to turn off the equipment.
Emergency Switch
Air Switch
Key switch