EXT TRIG POLARITY
Syntax:
POLAR Edge
POLAR?
Form:
Set & Query
Description:
Edge: Rising Edge or Falling Edge
Parameter:
Edge: R_E or F_E
RST value
-
Specified limit
R_E or F_E parameters
Example:
Falling Edge: POLAR F_E
Query mode: POLAR? :F_E
EXT TRIG PRESCALER
Syntax:
TPRESC Value
TPRESC??
Form:
Set & Query
Description:
Set prescaler value to divide Trig In source frequency
Parameter:
Value: Trig In prescaler value
RST value
-
Specified limit
from 1 to 2
16
-1
Example:
4: TPRESC 4
Query mode: TPRESC? :4
FREQUENCY
Syntax:
FREQ Fx, F
FREQ? Fx
Form:
Set & Query
Description:
Set internal mode frequency
Parameter:
Fx : F1, F2
F: frequency Hz
RST value
-
Specified limit
1 Hz to 1 MHz
Example:
FREQ F1, 1000
FREQ? F1=> :FREQ 1000
PERIOD
Syntax:
PERIOD Px, T
PERIOD? Px
Form:
Set & Query
Description:
Set internal mode period
Parameter:
Px : P1, P2
T: period in ns (5ns resolution)
RST value
-
Specified limit
1000ns to 4s
Example:
PERIOD P1, 1000
PERIOD? P1=> :PER 1000
3/2/18
Berkeley Nucleonics Corporation, 2955 Kerner Blvd. San Rafael, CA 94901
[email protected], Call 800-234-7858 or Live-Chat @ www.berkeleynucleonics.com