Service Manual for RESmart
®
Auto CPAP V1.1
11/16
12
Connect blower to connector and fix by clip. Put
the integrated one in the right place.
13
Covered by foam cover and clean up wires.
14
Assembly fixing clip on the main board PCBA.
15
Plug the 8-pin plug of blower in the socket PM
plug #2 as in Fig. E in the socket P24 on the
main board PCBA.
16
Assembly the main board PCBA on fixing pole
and make sure the clip clamps power PCBA.
17
1. As per Table 1, connect sensor YL to outlet
by silicon rubber canal.
2. Connect sensor LL to outlet by silicon rubber
canal.