F2 Full Face Mask User Manual V2.1

 

8

 / 

14

 

E

n

g

lis

h

 

Cleaning the Mask at Home   

Notes:

 

 

The  mask  and  headgear  can  only  be  cleaned  through  hand 
washing. 

 

The  elbow,  the swivel  and the  valve  cannot be dissembled  for 
washing. 

 

CAUTIONS

 

 

Do  not  use  solutions  containing  bleach,  chlorine,  alcohol, 
aromatics, moisturizers, antibacterial agents, or scented oils to 
clean any part of the system or air tubing. These solutions may 
cause damage and reduce the life of the product.   

 

Exposing any  part  of the  system  or tubing to direct sunlight  or 
heat may cause deterioration.   

 

If  any  visible  deterioration  of  a  component  is  apparent  (cracking, 
crazing,  tears  etc),  the  component  should  be  discarded  and 
replaced.   

Daily / After Each Use                                                                           

 

 

Disassemble  the  mask  components  according  to  the 
disassembly instructions. 

 

Thoroughly  clean  the  separated  mask  components  (excluding 
headgear),  by  gently  rubbing  in  warm  (approx.  30°C)  water 
using mild soap. 

 

Use a soft bristle brush to clean the vent. 

 

Rinse all components well with drinking quality water and allow 
to air dry out of direct sunlight. 

 

When  all  components  are  dry,  reassemble  according  to  the 
reassembly instructions. 

Summary of Contents for BMC-FM2

Page 1: ...English F2 Full Face Mask User Manual 0123...

Page 2: ...ask 2 Components of the F2 Full Face Mask 4 Getting the Right Cushion Size 5 Fitting the Mask 6 Disassembling the Mask 7 Cleaning the Mask at Home 8 Reprocessing the Mask between Patients 9 Reassembli...

Page 3: ...r mask Intended Use The F2 Full Face Mask channels airflow non invasively to a patient from a positive airway pressure device such as a continuous positive airway pressure CPAP or bi level system The...

Page 4: ...s warning applies to most models of CPAP systems At low CPAP pressures the flow through the exhaled port may be inadequate to clear all exhaled gas from the tube Some rebreathing may occur To minimize...

Page 5: ...generator If it is used beyond technical specifications or used with incompatible devices the seal and comfort of the mask may not be effective optimum therapy may not be achieved and leak or variati...

Page 6: ...F2 Full Face Mask User Manual V2 1 4 14 English Components of the F2 Full Face Mask Components of the F2 Full Face Mask Cushion Forehead pad Frame assembly Headgear Elbow assembly Elbow Valve Swivel...

Page 7: ...lowing drawing describes the different features of face and the length of the face H you need to measure The masks are available in three different sizes Choose the appropriate size according to the t...

Page 8: ...that the label on the back of the headgear faces outward when the headgear is worn Bring the lower half straps beneath the ear and encircle the face and then insert it into the fixed hook on the lower...

Page 9: ...ly cannot be disintegrated 3 The valve cannot be dissembled or damage may be incurred and the re installation will become more difficult 1 2 3 Undo and remove the upper half headgear straps out of the...

Page 10: ...g any part of the system or tubing to direct sunlight or heat may cause deterioration If any visible deterioration of a component is apparent cracking crazing tears etc the component should be discard...

Page 11: ...and disinfection instructions are available from the BMC website www bmc medical com If you do not have internet access please contact your BMC representative Reassembling the Mask 1 2 3 Install the...

Page 12: ...sure the headgear is not over tightened Mask size is wrong Consult your clinician Mask leaks around the face The cushion is misplaced on the cushion frame Check insertion of the cushion and reinsert c...

Page 13: ...sizes but is less than 218 mL Therapy Pressure 3 to 40 cmH2O Resistance Drop in Pressure measured average for 3 sizes at 50 L min 0 15 cmH2O at 100 L min 0 5 cmH2O Inspiratory and Expiratory Resistanc...

Page 14: ...0 dBA with uncertainty 3 dBA Environmental Conditions Operating temperature 5 C to 40 C 41 F to 104 F Operating humidity 0 93 relative humidity non condensing Storage and transport 20 C to 55 C 4 F to...

Page 15: ...nts Indicates a Warning or Caution and alerts you to a possible injury or explains special measures for the safe and effective use of the device Lot number Catalog number Not made with natural rubber...

Page 16: ...To exercise the rights under this warranty contact the local authorized dealers The expected service life of F2 Full Face Mask is one year MANUFACTURER BMC MEDICAL CO LTD Room 110 Tower A Fengyu Buil...