background image

 

AQS610-EU/SF  (V2 H/W) 

DOC 584 Issue A 

TECHNICAL GUIDE 

Page 20 

 

7.  Troubleshooting 
 

7.1  Inconsistent leaktest results 

Could be due to: 
1.  Bad neck sealing.  Check that the screw is positioning bottles directly under test heads and that sealing pads 

are undamaged. 

2.  Clamp  pressure  too  low.    Causes  poor  neck  seal  with  bottle  neck.    Increase  clamp  pressure  regulator 

slightly. 

3.  Clamp  pressure  too  high.    Causes  bottle  necks  to  distort,  releasing  air  pressure  inside.    Reduce  clamp 

pressure regulator slightly. 

4.  Leaking pneumatic fittings or pipes.  Check that joints between fittings are airtight and pneumatic pipes are 

not damaged. 

5.  Sense  pipe  blocked.    Pressure  readings  will  show  a  sluggish  response  to  pressure  changes.    Check  that 

sense  pipes  from  test  heads  to  pressure  transducers  on  main  control  board  are  not  blocked/kinked, 
especially when cabinet front panel is shut.  

6.  Fill time set too low. Bottle not reaching test pressure in allotted time. Increase fill time Setting. 
7.  Test  pressure  set  too  low.  Bottle  not  being  pressurised  enough  to  expose  very  small  holes.  Increase  test 

pressure Setting. 

8.  Fill  regulator  set  too  low.  Not  enough  air  flow  to  fill  the  bottle to test pressure in allotted time. Increase Fill 

regulator. 

9.  Test time too short. Not enough time allowed for a reliable Leaktest. Increase Test time Setting. 
 

7.2  Failure to reject bad bottles 

If the test results indicate that a bottle has been failed but was not rejected from the conveyor: 
1.  Check screw homing position sensor is operating correctly. 
2.  Check  reject  blower  OK.    Valve  can  be  forced  “ON”  from  the  front  panel  by  turning  on  the  function 

corresponding to the reject blower output to check the electronics and the valve solenoid.  Manually turning 
on the valve will check the valve operation. (See Functions section and I/O list and front panel controls for 
details). 

3.  Check that the ‘Reject Time’ setting is et such that the operation time matches the time taken for one screw 

revolution. 

 
If the test results do not indicate that a bottle should have been failed: 
1.  Check the ‘Final P%’ for the bottle tested.  Adjust the ‘Min Final P %’ setting if necessary to reject failures but 

to pass good containers. 

2.  Check transducer is not damaged or that pneumatics are connected and intact. 
3.  Fill regulator set too high. Bottles are over pressurising. Reduce Fill regulator. 
4.  Test  pressure  too  low.  Sometimes  a  higher  test  pressure  is  necessary  to  expose  smaller  holes.  Increase 

‘Test Pressure’ setting.  

5.  Fill Time too low. Bottle not reaching test pressure in allotted time. Increase ‘Fill Time’ setting. 
6.  Test time too short. Not enough time allowed for a noticeable pressure drop in the bottle (for  a small hole). 

Increase ‘Test Time’ Setting. 

 

7.3  Bottles not filling 

Occasionally a bottle may not fill correctly if the neck seals and bottle do not line up for any reason, e.g. bottle 
wobbling in test station too much .Other factors may cause the bottles not to fill on a regular basis: 
1.  Test head neck seal damaged. Replace faulty seal. 
2.  Test head bracket set too high. Cylinders do not reach bottles when extended.  Lower test head bracket. 
3.  Clamp pressure set too high. Bottles distorted and not sealing on test seals. Lower clamp pressure 

regulator. 

4.  Clamp pressure set too low. Head not making a good seal, air leaking around bottle neck. Increase clamp 

pressure regulator. 

5.  Fill  pressure  too  high.  Lower  test  pressure  (not  below  15mB  (  0.15  psi  )),  increase  fill  valve  pressure 

regulator or increase ‘Fill Time’ setting. 

Summary of Contents for AQS610-EU-SF

Page 1: ...5HZ The majority of equipment designed and manufactured by BMC Controls Ltd is protected by patents and industrial design copyright Examples of this include our PLASMA leak testing technique stepper m...

Page 2: ...TAL DISPLAY 12 4 2 KEYPAD 12 4 3 PAGE SELECT KEYS 12 4 4 CURSOR KEYS 12 5 DISPLAY PAGES 13 5 1 PARAMETERS 13 5 1 1 Test Information 13 5 1 2 Test Parameters 14 5 2 FUNCTIONS 15 5 2 1 I O Forcing 15 5...

Page 3: ...rejected as they leave the screw If the unit is for the U K and European market it uses mB as its pressure measurement unit the version for the USA uses Psi 1 1 Leaktest station The standard leaktest...

Page 4: ...e SCREW ERROR will be displayed 4 After a delay set by the operator Head Delay the Test head cylinders are extended this can be set to zero if no delay is required 5 The Fill valves are turned on 6 Th...

Page 5: ...tiated as the screw starts to turn to bring the next set of bottles into the tester 2 There is a settable delay Reject Delay which elapses before the reject routine proper starts to ensure that bottle...

Page 6: ...System layout drawing NOTE The reject position is the last position where the bottle is still held in position by the screw Bottle indexing screw Test heads Dead plate Direction Of Travel Start P S Le...

Page 7: ...L RATE regulators to zero Initially fully restrict the Reject Blower Adjust the guide rails so that bottles travel centrally down the conveyor 3 2 Proximity Sensor The screw home position detection se...

Page 8: ...esent the photocell with a typical background object that is to be ignored e g hold a bottle behind the far guide rail in line with the sense path 5 Press the button once and release The photoswitch i...

Page 9: ...pressure should be such that the test heads should be able to be lifted off the bottle necks by hand using a moderate amount of force such as that required to lift a bag of sugar with one hand Force...

Page 10: ...onds The test pressure Test Press is normally left set to 30 mB 0 5 Psi However this can be adjusted if necessary if there is a specified test pressure or if the bottle is being distorted To set the M...

Page 11: ...assed decrease Reject Delay NOTE When optimising the reject station settings very small changes should be made between tests i e not more than 0 02 seconds per change Repeat this procedure until clean...

Page 12: ...ill re start this 5 minute countdown When the backlight is not illuminated pressing any key on the AQS keypad will cause the back light to illuminate again and the 5 minute countdown to re start 4 2 K...

Page 13: ...d see troubleshooting or no bottle was present in that test position O The previous container failed the leaktest as over pressure some system adjustment is required see troubleshooting There are also...

Page 14: ...in channel 4 9 Pressure 5 LIVE pressure reading for channel 5 10 Final P 5 Final pressure of the last bottle leak tested in channel 5 11 Pressure 6 LIVE pressure reading for channel 6 12 Final P 6 Fi...

Page 15: ...display will show the currently selected input or output along with it s forced status and it s normal status Inputs and outputs are shown in groups of four the one to be altered will be highlighted w...

Page 16: ...may look as follows OP s 1000101011001 1 IP s 0010 10 Output 1 is normally on Input 5 is forced on Output 6 is normally off Output 16 is forced off IMPORTANT After I O forcing always ensure that all f...

Page 17: ...s 7 Reject On Time Time for which the Reject Blower activates to remove a failed bottle from the container this must be carefully set to match the screw revolution time 8 Reject Off Time Delay between...

Page 18: ...rrupted contact BMC Controls for a service visit Er 11 Parameter Checksum error some or all of the system parameters have become corrupted check and re enter the system parameters as described below i...

Page 19: ...eased until the optimum value was found which gave the highest speed yet maintained smooth operation this was found to be 2250 to convert this to an actual speed it must be divided by the gearbox rati...

Page 20: ...reject blower output to check the electronics and the valve solenoid Manually turning on the valve will check the valve operation See Functions section and I O list and front panel controls for detail...

Page 21: ...b99 13 set to 1 and Enable IP made on X7 pin 1 connected to pin 5 The screw does turn 1 If the screw rotates through more than 1 revolution at power up then stops with SCREW ERROR displayed The screw...

Page 22: ...ts are PNP i e switch 24V onto the signal line OUTPUT FUNCTION HARDWARE 1 HEAD_CLAMP SOLENOID VALVE CYLINDER 2 REJECT SOLENOID VALVE BLOWER 3 UNUSED 4 UNUSED 5 UNUSED 6 HOME_SCREW RELAY OP TO LVD DRIV...

Page 23: ...NB All inputs are PNP i e switch 24V onto the signal line LVD2 X7 Input Connector Pin Usage Pin No Function 5 24 V Common 1 Input 0 Enable 9 Input 1 2 Input 2 10 Input 3 3 Input 4 11 Input 5 4 Input...

Page 24: ...em i e already mounted on a conveyor an emergency stop button will already be fitted If the unit is supplied in kit form the emergency stop circuitry is still fitted in the cabinet but is linked out i...

Page 25: ...ICAL GUIDE Page 25 10 Control Panel connections INPUT 1 INPUT 2 INPUT 3 NOT USED PRESS SENSE1 NOT USED PRESS SENSE2 NOT USED CABLETO PNEUMATIC CABINET PRESS SENSE3 RS232 PORT PRESS SENSE4 INPUT EXPANS...

Page 26: ...c Parts 290 4651 TM24064BBCW CC XT3 CP CC XT3 8SIP CC XT3 16OP CC XT3 MB Chameleon for Analogue card for Topbox Connector LCD Display AQS Keypad AQS Input Card AQS OP Card AQS Motherboard Micro Proces...

Page 27: ...his test head is a solid construction with a flat mating face no flexibility Mechanical block bore This test head has a plunger which enters the neck of the bottle if the plunger is pushed a mechanica...

Reviews: