Troubleshooting
©
by
Blum
‐
Novotest
33
V2A
|
04/2018
Fig.
7
‐
1
Provide
pressure
compensation:
1) Untighten
the
screws
at
the
battery
compartment
lid.
2) Pull
back
both
screws
and
remove
the
battery
compartment
lid.
3) Loosen
the
setscrew
(at
least
with
2
revolutions).
4) Deflect
stylus
several
times
in
Z
‐
direction.
Wait
for
approx.1
minute.
5) Then
tighten
the
setscrew
again.
6) Put
in
the
battery
compartment
lid
precisely
and
mount
it
(M
d
= 0.7 Nm).
Thereby,
please
check
the
O
‐
rings
and
grease
it,
if
necessary.
Error
LED
Status
after
battery
change
(40) Setscrew
Error
LED
Status
flashes
green
‐
blue
resp.
green
‐
orange
Probe
system:
LED
Status
flashes
green
‐
blue
resp.
green
‐
orange.
IR
receiver:
LED
Receive
IR
off.
Probe
does
not
respond
to
M
‐
Code
(switch
‐
on).
Check
wiring
and
switch
‐
on
command
on
the
IF59:
(IF59
X1:
Pin
3,
Pin
4,
Pin
5)
Check
programming
of
the
probe
(Mode
TC
‐
Mode
DUO)
Summary of Contents for TC50