background image

 

BlueEyes Technology 

www.BlueEyes.com.tw

 

 

2. Specification 

Sensor 

Uncooled focal plane 

Temperature range 

30˚C~45˚C 

Resolution 

0.1˚C 

Accuracy 

±0.5˚C @1m 

Minimum measuring distance 

15cm 

Temperature  measurement  response 
time 

≤500ms (95% of reading) 

Pixel of thermal imagery 

19200 (160×120) 

Pixel dimensions 

12μm 

Palettes 

Iron red, rainbow, white, black, red, lava, high-contrast rainbow 

Bandwidth of infrared spectroscopy 

8~14μm 

Field of view (FOV) 

56°(H) × 42°(V) 

Instantaneous field of view (IFOV) 

6mrad 

Thermal imaging sensitivity 

50mk 

Frame rate 

9Hz 

Temperature display 

Center  temperature  measurement  and  high  temperature 
tracking (default) 

Image format 

BMP 

Image modes 

Thermal imagery, digital camera (visible light), fusion 

Temperature measuring point 

Besides  central  point,  3  temperature  measuring  points  can  be 
added 

Visible light 

√ 

Resolution of visible light 

640×480 

Display resolution 

320×240 

Mixed setting 

0%  (total  visible  light),  25%,  50%,  75%,  100%  (total  thermal 
imagery) 

PC analysis software (PC) 

√ 

Real-time image transmission 

√ (PC software image projection) 

Data communication 

Type-C USB 

Auto power off 

Selectable  (5min,  10min,  30min),  30min  auto  power  off 
(default) 

Service time 

6 hours 

Charging time 

5 hours 

Charging voltage/current 

5V/2A 

Image storage 

Micro SD card 

Power 

A single Li-ion 3.7V/5000mAh 26650 battery 

Display 

2.8" TFT LCD 

Product size 

236mm x 75.5mm x 86mm 

Standard accessories 

Manual, Type-C USB cable, 16GB TF card 

Standard carton gross weight 

2.2kg 

Summary of Contents for BE-5302

Page 1: ...Thermal imager User Manual 2020 4 24 www BlueEyes com tw...

Page 2: ...without notice BlueEyes Technology is under no obligation to assume any responsibility for any errors in the information contained in this manual including software firmware and hardware Any product...

Page 3: ...rmation 12 2 Specification 13 3 Appearance 14 4 Show Description 15 5 Power on off 16 6 Calibrating 16 7 Measurement 17 8 Palette 18 9 Point Temperature 20 10 Image Mode 21 11 Settings 23 11 1 Languag...

Page 4: ...nology www BlueEyes com tw 19 Maintenance 41 20 USB connection to PC 42 21 Safety 43 22 Notes for use 43 23 Software 44 23 1 System Requirements 44 23 2 USB mode for software 44 23 3 USB camera 45 23...

Page 5: ...e sure to make sure the version you use is the same as the one in this manual 4 Please ask us again for the latest version of the manual if it is different 5 If you want to check the current firmware...

Page 6: ...electric shock or product damage do not expose the machine to rain dampness or liquid splashes and do not place water containers such as vases on the machine 4 Do not place this machine in a bookcase...

Page 7: ...uch as fans air conditioning outlets etc 16 When the measurement object comes from a place where the temperature difference in the measurement environment is relatively large it is necessary to stay i...

Page 8: ...to the emissivity and reflection of the object s apparent temperature parameters 26 The application of infrared thermal imaging cameras has its limits In fact it is judged by experienced users that t...

Page 9: ...the use of railings or red dragon posts to guide the queue of people the measurement of the moving line to a single person to pass slowly as the main requiring the subject to take off the hat do not...

Page 10: ...other institutions depending on the average height of the students can be adjusted about 10 15 cm or so above the average height infrared video thermometer shooting angle to look down 10 15 degrees f...

Page 11: ...r areas The temporal artery from the forehead to the upper edge of the temporal cavity is the main temperature detection area which may affect the accuracy of the temperature measurement if the subjec...

Page 12: ...ck may occur or damage may be caused to the machine 3 When the machine is used in an abnormal way this generally means that it needs to be repaired 4 When replacement parts are required ensure that th...

Page 13: ...o need to calibrate the temperature of this equipment In the event of a large measurement deviation and the automatic calibration is not possible you can send it back to us for calibration 1 12 Update...

Page 14: ...ult Image format BMP Image modes Thermal imagery digital camera visible light fusion Temperature measuring point Besides central point 3 temperature measuring points can be added Visible light Resolut...

Page 15: ...urn white light on or off 3 SD card slot 13 Left button 4 Lower casing 14 down key 5 Terminal Protection Cover 15 up key 6 White Lighting 16 Photo Browser 7 Infrared Thermal Imaging Lens 17 right clic...

Page 16: ...tion No Description 1 Center temperature 6 High Temperature Warning 2 Automatic tracking of temperature maximums 7 Automatic tracking of temperature maximums 3 Auto calibration in progress 8 Temperatu...

Page 17: ...e to the protection mechanism of the device The thermal imaging camera needs to go through 5 6 seconds after starting the shutdown procedure Please try to avoid repeatedly switching the machine on and...

Page 18: ...the main menu 2 Press the left right button to select the measurement option 3 Press the SET button to enter the measurement selection menu 4 Press the up down button to select whether the center or...

Page 19: ...Lava Rainbow HC etc The BE 5302 is available in a variety of color palettes with the following steps 1 On the Thermal Imaging Measurement screen press SET to call up the main menu 2 Press the left ri...

Page 20: ...BlueEyes Technology www BlueEyes com tw Iron Rainbow White Hot Red Hot Black Hot Lava Rainbow HC...

Page 21: ...ess the up and down buttons to select different temperature measurement points to turn on or off 4 Select the current temperature measurement point and press SET to enter the thermal imaging measureme...

Page 22: ...the main menu 2 Press the left right button to select the photo mode configuration options 3 Press the SET button to enter the photo mode selection menu 4 Press the up down button to select different...

Page 23: ...al Digital Fusion Press the Back button to return to the Main Menu and Thermal Imaging In the fusion imaging mode after returning to the thermal imaging measurement screen you can select different fus...

Page 24: ...o select different setting options 5 Press the SET key to enter the corresponding setting options and configure the parameters 6 Press the Return button to return to the previous menu level 11 1 Langu...

Page 25: ...nu press the up down button to select the date and time setting option 2 Press the SET key to enter the date and time setting 3 Select different time nodes and time formats by left right button 4 Pres...

Page 26: ...ng the setting menu press the up and down button to select the temperature unit setting option 2 Press the SET key to enter the temperature unit setting 3 Press the up down button to select different...

Page 27: ...ning setting option 3 Press the SET key to enter the high low temperature warning setting 4 Press the up down button to select different settings to turn on or off the high low temperature warning and...

Page 28: ...BlueEyes Technology www BlueEyes com tw When the warning is turned on an alarm will appear in the screen when the temperature exceeds the set range as follows...

Page 29: ...key to enter the measurement parameter setting 3 Press the up and down button to select the emitting rate or temperature compensation 4 Press the SET button to enter the setting 5 Press the up down b...

Page 30: ...tton to select the brightness setting option 2 Press the SET key to enter the display brightness setting 3 Press the up down button to select different display brightness levels 4 Press the SET key to...

Page 31: ...t the Auto Off setting option 2 Press the SET button to enter the auto shutdown setting 3 Press the up and down button to select a different auto off time or to turn off the auto off 4 Press the SET k...

Page 32: ...mode setting 3 Press the up down button to select a different USB mode 4 Press the SET key to select and save the settings 5 Press the Return button to return to the previous menu level There are two...

Page 33: ...rage settings 3 Press the up and down button to select whether or not to turn on the automatic storage function 4 Press the SET key to select and save the settings 5 Press the Return button to return...

Page 34: ...etup menu press the up down button to select the system setup options 2 Press the SET button to enter the system settings menu 3 Press the up and down button to select the corresponding system setting...

Page 35: ...y following the steps 1 In the system settings press the up down button to select the device information option 2 Press SET button to access the device information 3 Press the Return button to return...

Page 36: ...ress the up down button to select the Restore Factory Settings option 2 Press the SET key to enter then press the up down button to select whether to resume the factory setting 3 Press SET key to conf...

Page 37: ...gs press the up down button to select the format SD card option 2 Press the SET key to enter 3 Press the up down button and press the SET key to select whether to format the SD card 4 Press the Return...

Page 38: ...of other objects is between 0 and 1 Emission rate is a pure quantity In general the darker the color of the material or the rougher the surface the closer the emissivity will be to 1 The following is...

Page 39: ...gy www BlueEyes com tw 13 Temperature compensation Temp comp To adapt to different environments and venues you can perform manual temperature compensation Please see 10 5 measurement parameters for de...

Page 40: ...n to enter the Photo Viewer screen 2 Select different photos via the up down left right buttons 3 Press the SET key to access the full screen In full screen mode 4 Tap the SET button to access the pho...

Page 41: ...n 3 Press the SET key to confirm or press the Return key to return to cancel this photo Photos are automatically saved to an SD card in auto storage mode Photos can only be saved when an SD card is in...

Page 42: ...ing speed please back up the data regularly and clean the data on the card in a timely manner To avoid SD card data abnormalities caused by hot disconnect please do not disconnect the SD card repeated...

Page 43: ...k mode connect the USB cable to your computer and use the software to browse photos and analyze data 3 After setting the USB mode to USB camera mode connect the USB cable to the computer and use the s...

Page 44: ...he actual temperature otherwise the measured temperature will be inaccurate To ensure the accuracy of the product it is recommended to preheat the machine for 10 to 15 minutes before measurement when...

Page 45: ...re supports the following operating systems Microsoft Windows 7 Microsoft Windows 8 Microsoft Windows 10 If the operating system is lower than the above version please update the operating system firs...

Page 46: ...ctions Zone Description 1 Option to view live image or alarm history 2 Live Image 3 alarm setting 4 Alarm Information 5 List of alarm history Full screen mode Press Zone 2 twice to enter full screen m...

Page 47: ...he maximum temperature exceeds or equals the alarm valve value When the alarm function is enabled the maximum temperature value in zone 4 is displayed as a warning color flashing When an audible alarm...

Page 48: ...e met Change the alarm file For example you can add a guiding phrase such as If you are passing through the test area please go to the review area for further temperature measurement thank you for you...

Page 49: ...BlueEyes Technology www BlueEyes com tw 23 4 USB disk Area Description 1 Menu 2 Option to view a list of images or image information 3 List of folders in SD memory card 4 Picture List 4 3 2 1...

Page 50: ...com tw View image Double clicking on any of the 4 image files in the Area 4 image list will open the current image file and display the relevant information The details of the picture will be displaye...

Page 51: ...ools Used to select an existing drawing the selected drawing is shown as a dotted line Select points lines rectangles and circles to draw on the picture and on the right side you will record the data...

Page 52: ...ge files Other In the bottom right corner of the window the temperature of the Meta point where the image cursor is located is displayed The top right corner of the window is used to set the temperatu...

Page 53: ...BlueEyes Technology www BlueEyes com tw Note...

Page 54: ...BlueEyes Technology www BlueEyes com tw Note...

Page 55: ...BlueEyes Technology www BlueEyes com tw Note...

Page 56: ...for informational purposes only BlueEyes is registered trademarks or trademarks of BlueEyes Technology Corporation Other product or company names mentioned herein may be the trademarks of their respe...