Blue-White ProSeries SONIC-PRO S4 Series Installation And Operating Manual Download Page 14

Page 14

Sonic-Pro

6.3.1
Logging
Menu

Data logs are stored in the meter’s internal memory in a space delimited .TXT file that can 
easily be copied to a removable USB flash drive located on the main circuit board. Each log 
file includes the date and time, the flow rate (FR) and the flow total (TL). A total of 191,800 
logs can be stored.

2013/12/29 09:15:00 FR 0.0000 TL 0.00

The 

Log Interval 

determines the number of seconds from 1 - 999,999 that the meter will wait 

between logs. 

If a value is entered into 

Rate Set Point 1,

 a new log will be generated every 60 seconds 

while the measured rate of flow is greater than the configured value.  

If a value is entered into 

Rate Set Point 2,

 a new log will be generated every 60 seconds 

while the measured rate of flow is less than the configured value.  

If a value is entered into any 

Tot. Set Point (1, 2, 3, 4 or 5),

 a new log will be generated 

when the total accumulated flow value is equal to the configured value.

To access the log entries, you must copy the logs onto a USB flash drive and transfer the file 
to a computer for reading. To copy all of the log entries, open the enclosure and insert a USB 
flash drive. Highlight 

Copy Logs 

press either right arrow 

u

 or 

ENTER. 

When copying logs, 

all of the logs must be copied. You may not copy only selected logs. 

To delete ALL of the log entries, highlight 

Delete Logs 

then press either right arrow 

u

 or 

ENTER. 

You may not delete only selected logs. All logs must be deleted.

Logging

Log Interval(s)

0.0000

Enter Value

Copy Logs

Delete Logs

Rate Set Point 1

0.0000

Enter Value

0.0000

Enter Value

Tot. Set Point 1

0.0000

Enter Value

0.0000

Enter Value

Rate Set Point 2

Tot. Set Point 2

0.0000

Enter Value

0.0000

Enter Value

Tot. Set Point 5

0.0000

Enter Value

Tot. Set Point 3

Tot. Set Point 4

Page 13

Sonic-Pro

6.3
Configure the
SPU Meter
Settings

In the 

Meter Settings 

menu, the data logging functions and the display features can be 

configured. 

Meter Settings

Logging

Log Interval(s)

0.0000

Enter Value

Copy Logs

Delete Logs

Display

Flow Rate

Volume Units

Ounces

U.S. Gallons

U.S. Barrels Liq.

U.S. Barrels Oil

Cubic Feet

Acre Feet

Imperial Gallons

Cubic Meters

Liters

Milliliters

FR Custom V/Gal

Time Units

Seconds

Hours

Days

Minutes

Fl. Rate Decimal

0

Enter Value

Totalizer

Volume Units

Ounces

U.S. Gallons

U.S. Barrels Liq.

U.S. Barrels Oil

Cubic Feet

Acre Feet

Imperial Gallons

Cubic Meters

Liters

Milliliters

FR Custom V/Gal

Fl. Total Decimal

0

Enter Value

Flow Average(s)

0

Enter Value

Enable Tot. Clear?

Yes

No

Low Fl. Cut-Off

0

Enter Value

High Fl. Cut-Off

999

Enter Value

Scaling Off-set

1.0000

Enter Value

Zero Off-set

Sign

FR Offset

0.0000

Enter Value

Negative

Positive

SOS Fault Window

10

Enter Value

Rate Set Point 1

0.0000

Enter Value

Sleep Time(s)

 0

Enter Value

0.0000

Enter Value

Tot. Set Point 1

0.0000

Enter Value

0.0000

Enter Value

Rate Set Point 2

Tot. Set Point 2

0.0000

Enter Value

0.0000

Enter Value

Tot. Set Point 5

0.0000

Enter Value

Tot. Set Point 3

Tot. Set Point 4

0

Enter Value

0

Enter Value

Industries, Ltd.

Industries, Ltd.

ProSeries

by Blue-White Ind.

d

Industries, Ltd.

Industries, Ltd.

ProSeries

by Blue-White Ind.

d

Summary of Contents for ProSeries SONIC-PRO S4 Series

Page 1: ...Installation and Operating Manual 2 TWO YEAR WARRANTY ProSeries Series S4 Ultrasonic Flow Meter Industries Ltd Industries Ltd ProSeries by Blue White Ind d SONIC PRO d...

Page 2: ...DING THE SAFE USE OF THIS PRODUCT AND OTHER TECHNICAL ASSISTANCE MAY BE DIRECTED TO Blue White Industries 714 893 8529 techsupport blue white com CAUTION risk of electric shock CAUTION risk of danger...

Page 3: ...ions of other chemicals or other chemicals entirely may be acceptable provided the chemical does not change the sound speed SOS more than approximately 100 meters per second and provided the fluid all...

Page 4: ...must not be cut or modified Note that the Sonic Pro can accurately measure flow from either direction The SPU can be mounted on a wall or on a horizontal or vertical run of pipe Although the Sonic Pro...

Page 5: ...is required on downward flows to ensure a full pipe See the minimum straight length of pipe requirement chart above Related piping must be independently supported The meter should not be allowed to s...

Page 6: ...d tight cable gland connectors for cable diameter from 200 to 394 inches 5 1 to 10 0 mm Two liquid tight cable gland connectors cable for diameters from 118 to 255 inches 3 0 to 6 5 mm One communicati...

Page 7: ...ector T2 1 2 Positive Negative Type Open collector w internal 1 K pull up to 5V Can be pulled up to 30V 10mA 3K pull up Duty Cycle 50 5 Frequency Range 0 1000 Hz Resolution 1 Hz Accuracy 0 5 F S PADDL...

Page 8: ...allowable SOS Speed Of Sound error set point has been exceeded FAULT 21 The configuration data is not valid Flow Direction Total Units U S Gallons Min 105 1482 25 05 469 U S Gallons 0 28 Transducer S...

Page 9: ...he value Press ENTER to confirm the modifications and return to the menu heading Press ESC to cancel the modifications and return to the menu heading 5 3 2 Selecting Optional Items Press up arrow p or...

Page 10: ...use your inlet outlet pipe material 10 Fluid Type Select WATER or CUSTOM 11 Fluid SOS If WATER is selected it is not necessary to input a fluid SOS If CUSTOM fluid type is selected the Speed of Sound...

Page 11: ...e factory set inside the pipe fitting and cannot be moved Select the mount method that is shown on the Calibration Detail label affixed to the pipe fitting Transducer Model I C J Mount Method W V Cabl...

Page 12: ...ings Direction Zero A B Scaling A B Enter Value Sign A B Zero Value A B Positive B A A B Enter Value Zero B A Scaling B A Enter Value Sign B A Zero Value B A Enter Value 0 975 0 455 Negative Positive...

Page 13: ...n the acceptable range and there are not too many particles or other sound blocking or reflecting contaminants in the fluid which can inhibit the sound beam The target Fluid SOS value can be adjusted...

Page 14: ...els Oil Cubic Feet Acre Feet Imperial Gallons Cubic Meters Liters Milliliters FR Custom V Gal Fl Total Decimal 0 Enter Value Flow Average s 0 Enter Value Enable Tot Clear Yes No Low Fl Cut Off 0 Enter...

Page 15: ...into any Tot Set Point 1 2 3 4 or 5 a new log will be generated when the total accumulated flow value is equal to the configured value To access the log entries you must copy the logs onto a USB flash...

Page 16: ...ters Milliliters TOT Custom V Gal Fl Total Decimal 0 Enter Value Flow Average s 0 Enter Value Enable Tot Clear Yes No Low Fl Cut Off 0 Enter Value High Fl Cut Off 999 Enter Value Scaling Off set 1 000...

Page 17: ...Off value totalizing will still occur 6 3 2 6 High Flow Rate Cut off Setting High Fl Cut Off 999 Enter Value Scaling Off Set is a multiplier used to correct for flow inaccuracies All flow rate values...

Page 18: ...e meter was sleeping using the following formula Total flow display value T1 F1 F2 2 x N 1 Where T1 Last total before sleep F1 Last flow rate units per second before sleep F2 Flow rate average of the...

Page 19: ...gh Release A 0 Enter Value Low Trigger A 0 Enter Value Low Release A 0 Enter Value Time s A 0 Enter Value Monitor Totalizer Batch Amt A 0 Enter Value Contact Timer s A 0 Enter Value Faults A Relay B M...

Page 20: ...ue Rate at Max Curr 0 0000 Enter Value Max Analog O P 20 Enter Value 6 4 2 Pulse 0 1000 Hz Output Menu A digital pulse output signal can be programmed within the range of 0 1000 Hz pulses per second S...

Page 21: ...ue Faults B None A None B Relay B Monitor Flow Rate High Trigger B 0 Enter Value High Release B 0 Enter Value Low Trigger B 0 Enter Value Low Release B 0 Enter Value Time s B 0 Enter Value None B 6 4...

Page 22: ...Value Contact Timer s B 0 Enter Value 6 4 3 2 CC Assigned to Monitor FAULTS When in the Relay A or Relay B menu selecting and then exiting the FAULTS menu will assign that relay to energize when any...

Page 23: ...unsigned char 8 Frequency output 45 Firmware Version unsigned char 8 8 ASCII characters A11 22 33 53 Log Entry unsigned char 86 Log Entry String Eg 2013 12 29 09 15 00 FR 0 0000 TL 0 00 Description P...

Page 24: ...e meter can be re calibrated in the field by selecting Calibration and pressing the right arrow u or ENTER button The pipe must be full of fluid and there must be no movement of the fluid in the pipe...

Page 25: ...Save Config function is possible at the main menu Global Config Password 1234 Enter Value Change Password Change Password Old Pass 1234 OK CANCEL NEW Pass 5678 OK Enter master password 3408 to clear a...

Page 26: ...me Units Independently configurable Rate and Total display units in U S Gallons ounces barrels US liquid barrels US oil cubic ft acre ft Imperial British gallons liter cubic meter or user defined cust...

Page 27: ...ipe fitting transducer no transducer cables Standard 4 20mA and High Speed Pulse output only Without power supply customer supplies 15 30 VDC 0 25 ft cable length with field replaceable liquid tight c...

Page 28: ...03 5 41 1427 10 50 1447 20 68 1481 30 86 1507 40 104 1526 50 122 1541 60 140 1552 70 158 1555 80 176 1555 90 194 1550 100 212 1543 Water Sound Speeds 7 2 Sound speed data Page 27 Sonic Pro Industries...

Page 29: ...in 18 39cm 5 90in 14 97cm 74in 1 88cm 2 75in 6 99cm 7 24in 18 39cm 5 83in 14 80cm 55in 1 39cm 2 75in 6 99cm Sonic Pro Inline Pipe Fittings Pipe Size 2 1 5 1 75 5 Pipe Size OL OW A B C 2 7 47in 18 97cm...

Page 30: ...ers TOT Custom V Gal Fl Total Decimal 0 Enter Value Flow Average s 0 Enter Value Enable Tot Clear Yes No Low Fl Cut Off 0 Enter Value High Fl Cut Off 999 Enter Value Scaling Off set 1 0000 Enter Value...

Page 31: ...Enter Value Time s B 0 Enter Value Monitor Totalizer Batch Amt B 0 Enter Value Contact Timer s B 0 Enter Value Faults B Clock CLOCK SET 2013 07 24 14 31 OK CANCEL Calibration Clear Cal Value None A N...

Page 32: ...kmanship for 24 months from date of factory shipment WARRANTY COVERAGE IS LIMITED TO REPAIR OR REPLACEMENT OF THE DEFECTIVE FLOWMETER ONLY UNDER NO CIRCUMSTANCES SHALL BLUE WHITE BE LIABLE FOR ANY CON...

Reviews: