OPERATION

PUMP

Adapt Pro 

Elite™ 

Operating, Maintenance & Trouble Shooting Guide

• Place the mattress on top of the bed frame. 

• Secure the mattress onto the bed frame by using the elastic tie down straps.

• Ensure the CPR and BOLSTER RELEASE valve is in closed position before turning on the power.

• Position the Pump by its spring loaded hanger brackets over footboard of the bed.

• Take black hose connector and firmly push into position on the pump. A “clicking” sound will secure the connection.

• Plug in black Fowler Sensor Cable. Be careful to ensure it clicks into place.

• Connect the power cord to the Pump and Press the red protector downward to secure the power cord.

• For patient transport, press “Auto Firm” button and wait until the mattress to be inflated. Disconnect the

hose from the Pump unit and put on the hose connector Transport Cap to keep the mattress inflated.  

• Push “Mode” button for Alternation/Alternating Air Therapy Mode. Select Cycle Time: 10-30 minutes per cycle. 

• For PST Auto-detection hold soft-firm button until light blinks. In approx 10 mins. pump will set based on patient’s weight.

• Two Static Modes: Static Care - After 20 mins. will fall back to alternation, Static - remains static until setting is changed.

• To lock in any mode, press lock-out button.

• To release, hold lock out button until light disappears.

• If not using PST, select comfort setting based on average patient weight.

In emergency: Twist CPR and BOLSTER RELEASE Valve in either direction to release air from mattress.

CPR 

Bolster Release

OPERATION

PUMP

Adapt Pro 

Elite™ 

Operating, Maintenance & Trouble Shooting Guide

• Place the mattress on top of the bed frame. 

• Secure the mattress onto the bed frame by using the elastic tie down straps.

• Ensure the CPR and BOLSTER RELEASE valve is in closed position before turning on the power.

• Position the Pump by its spring loaded hanger brackets over footboard of the bed.

• Take black hose connector and firmly push into position on the pump. A “clicking” sound will secure the connection.

• Plug in black Fowler Sensor Cable. Be careful to ensure it clicks into place.

• Connect the power cord to the Pump and Press the red protector downward to secure the power cord.

• For patient transport, press “Auto Firm” button and wait until the mattress to be inflated. Disconnect the

hose from the Pump unit and put on the hose connector Transport Cap to keep the mattress inflated.  

• Push “Mode” button for Alternation/Alternating Air Therapy Mode. Select Cycle Time: 10-30 minutes per cycle. 

• For PST Auto-detection hold soft-firm button until light blinks. In approx 10 mins. pump will set based on patient’s weight.

• Two Static Modes: Static Care - After 20 mins. will fall back to alternation, Static - remains static until setting is changed.

• To lock in any mode, press lock-out button.

• To release, hold lock out button until light disappears.

• If not using PST, select comfort setting based on average patient weight.

In emergency: Twist CPR and BOLSTER RELEASE Valve in either direction to release air from mattress.