Load Connection
27
Warning:
Each terminal can carry up to 30 A current. Double-terminal connection is needed when the input current
is more than 30 A. Refer to Figure
- Connection of Load and Device Under Test (DUT) for double-
terminal connection.
3.1 Parallel Connection
Parallel connection can be applied between modules of the same model to increase current and power dissipation, but
it
cannot be applied between different modules
.
Modules can be paralleled in CC, CR, or CW mode, but
cannot be paralleled in CV or CZ mode
.
Each module will dissipate the power it has been programmed at. For example, after being paralleled, two single-channel
modules rated at 80V/40A/300W can dissipate up to 80V/80A/600W.
Figure
illustrates the paralleled connection for increased power dissipation.
Figure 3.2
Parallel Module Connection