Enter a different current value, wait until the instrument is in meter mode, then make sure that the
displayed current value (actual output value) is the same as the set value.
12)
Ensure that the current can be adjusted from zero to the maximum rated value.
13)
Turn off the power supply and remove the short wire from the output terminals.
14)
Verify channels 2 & 3 according to steps 1) – 8)
5
.
If the Power Supply Does Not Turn On
Use the following steps to help resolve problems you might encounter when turning on the instrument.
1). Verify that there is AC power applied to the power supply.
Verify that the power cord is firmly plugged into the power receptacle on the rear panel of the power supply.
Make sure the power outlet you are using is working properly and verify that the power supply is turned on.
2). Verify the power-line voltage setting.
Make sure the voltage selector switch is set according to the present line voltage (110VAC or 220VAC).
Change the voltage setting if it’s not correct.
3). Verify that the correct power-line fuse is installed.
Model
Fuse Description
9130
Fuse 3.15A T250V(220V AC)
Fuse 6.30A T250V(110V AC)
If the fuse is blown replace it according to the table above.
6
.
To Adjust the Carrying Handle
To adjust the position, grasp the handle on each side and pull outwards, then rotate the handle to the desired
position.