CHAPTER 3. FRONT PANEL OPERATION
15
Figure 3.1: Measurement Timing
Figure 3.2: Measurement Timing 2
The data updating cycle refers to the cycle used for calculating
sampling data of the measurement function. It is identical to the set
value of the data updating rate.
The slope refers to signal changes from low level to high level (rising edge)
or from high level to low level (descending edge).
Measurement interval for harmonic measurement
At the sampling frequency of harmonic measurement, the measurement interval refers to the first point 1024 from
the data updating cycle. The sampling frequency of harmonic measurement is determined automatically by the
signal cycle set as the PLL source in the instrument. The sampling data or measurement interval for calculation may
be different from the sampling data or measurement interval of the measurement function in conventional
measurement.
3.4
Filter and Crest Factor Setting
1.
Select Menu
→
SET UP
→
OTHER SET and enter the OTHER configuration page.
2.
Press
to select the parameter to be configured (blue font background), and then press the soft key
corresponding to the parameter on the right to set the required value, as shown in the Figure ??.
3.
Press the Enter key to save the settings.