29
If x = I the measured frequency will be sent to the terminal.
Example:
Case 1
CFS4000 – sets the start frequency at 4000 Hz.
Case 2
CFU – increases the gate period used in measuring frequency
Message returned to terminal:
Case 1
:
“Output frequency: yyyyyyyy,yy Hz”
if the command was correctly sent.
“Wrong value sent”
if the “yyyyyyyyy”
parameter sent was not a valid value.
Case 2
:
“Gate period 0.1 s”
and “
Measured frequency: yyyyyyyyy Hz”
if the gate period is
changed.
“
Measured frequency: yyyyyyyyy Hz”
if asked about measured frequency.
CL command
Syntax:
CLx [yyyyy]
Parameters:
x represents a digit that has 2 valid values: “I” or “S”.
yyyyy is a decimal digit value between 10 and 10000, valid if x = S.
Description:
These commands set/interrogates the output level on the
OUTPUT
jack. The
command is available only in
EXTERNAL VCG
or
INTERNAL SWEEP
mode.
If x = S the output level will be set at the value of the “yyyyy”, a level that can take
values between 10mV and 10000mV.
If x = I the output level at the
OUTPUT
jack will be sent to terminal.
Example:
CLS345 – sets the output level at 345mV.
Message returned to terminal:
“Output level: yyyyy mV”
if the command was correctly sent
“Wrong value sent”
if “yyyyy” parameter sent was not a valid value.
CO command
Syntax:
COx [zyyyy]
Parameters:
x represents a digit that has 2 valid values: “I” or “S”.
z is a valid digit only if x = S and has 2 valid values: ”+” or “-“.
yyyy is a decimal digit value between 0 and 5000, valid if x = S.
Description:
This command sets/interrogates the offset level added at the
OUTPUT
jack. The
command is available only in
EXTERNAL VCG
and
INTERNAL SWEEP
mode.
If x = S, the offset level will be set at the value of the “zyyyy” parameter, offset that
can take values between –5000 mV and +5000 mV. It is necessary that the sign is
sent first; otherwise the following parameter will be dropped.
If x = I, the offset value at the
OUTPUT
jack will be sent to the terminal.
Example:
COS-4333 – sets the
OUTPUT
offset level at -4333 mV.
Message returned to terminal:
“Offset: zyyyy mV”
if the command is sent correctly.
“Wrong value sent”
if ”yyyy” parameter sent was not a valid value.