BK Precision 2522B Instruction Manual Download Page 20

MAINTENANCE

WARNING

The following instructions are for use by
qualified service personnel only. To avoid
electrical shock, do not perform any serv-
icing other than contained in the operat-
ing instructions unless you are qualified
to do so.

High voltage up to 2000 V is present when
covers are removed and the unit is oper-
ating. Remember that high voltage may
be retained indefinitely on high voltage
capacitors. Also remember that ac line
voltage is present on line voltage input
circuits any time the instrument is
plugged into an ac outlet, even if turned
off. Unplug the oscilloscope and dis-
charge high voltage capacitors  before
performing service procedures.

FUSE REPLACEMENT

If the fuse blows, the “ON” indicator will not light and the

oscilloscope will not operate. The fuse should not normally
open unless a problem has developed in the unit. Try to
determine and correct the cause of the blown fuse, then
replace only with the correct value fuse. For 110/125 V line
voltage operation, use an 800 mA, 250 V fuse. For
220/240 V line voltage operation, use a 600 mA, 250 V fuse.
The fuse is located on the rear panel adjacent to the power
cord receptacle.

Remove the fuseholder assembly as follows:

1.

Unplug the power cord from rear of scope.

2.

Insert a small screwdriver in fuseholder slot (located

between fuseholder and receptacle). Pry fuseholder
away from receptacle.

3.

When reinstalling fuseholder, be sure that the fuse is

installed so that the correct line voltage is selected (see
LINE VOLTAGE SELECTION).

LINE VOLTAGE SELECTION

To select the desired line voltage, simply insert the fuse

and fuse holder so that the appropriate voltage is pointed to
by the arrow. Be sure to use the proper value fuse (see label
on rear panel).

PERIODIC ADJUSTMENTS

Probe compensation and trace rotation adjustments

should be checked periodically and adjusted if required.
These procedures are given below.

Probe Compensation

1.

Connect probes to CH 1 and CH 2 input jacks. Per-

form procedure for each probe, one probe at a time.

2.

Set the probe to X10 (compensation adjustment is not

possible in the X1 position).

3.

Touch tip of probe to CAL terminal.

4.

Adjust oscilloscope controls to display 3 or 4 cycles of

CAL square wave at 5 or 6 divisions amplitude.

5.

Adjust compensation trimmer on probe for optimum

square wave (minimum overshoot, rounding off, and
tilt). Refer to Fig. 5.

Trace Rotation Adjustment

1.

Set oscilloscope controls for a single trace display in

CH 1 mode, and with the channel 1 AC-GND-DC
switch set to GND.

2.

Use the channel 1 POSition control to position the

trace over the center horizontal line on the graticule
scale. The trace should be exactly parallel with the
horizontal line.

3.

Use the TRACE ROTATION adjustment on the front

panel to eliminate any trace tilt.

Correct

Compensation

Over

Compensation

Insufficient

Compensation

Fig. 5.

Probe Compensation Adjustment.

18

Summary of Contents for 2522B

Page 1: ...20 MHz DIGITAL STORAGE ANALOG OSCILLOSCOPE BK PRECISION INSTRUCTION MANUAL MODEL 2522B BK PRECISION...

Page 2: ...wet 4 Use the time proven one hand in the pocket technique while handling an instrument probe Be particularly careful to avoid contacting a nearby metal object that could provide a good ground return...

Page 3: ...further precautionary information This symbol appears in the manual where the corresponding information is given Instruction Manual for Model 2522B 20 MHz Digital Storage Analog Oscilloscope 2000 B K...

Page 4: ...Trace Display 12 Triggering 13 Magnified Sweep Operation 15 X Y Operation 15 Video Signal Observation 15 OPERATING INSTRUCTIONS Continued Digital Storage Operation 15 Applications Guidebook 17 MAINTEN...

Page 5: ...otter CRT FEATURES Rectangular CRT Rectangular CRT with large 8 x 10 centimeter viewing area Convenience Trace rotation electrically adjustable from front panel 0 10 90 and 100 markers for rise time m...

Page 6: ...ng composite video waveforms TV H position can also be used as low frequency reject and TV V position can be used as high frequency reject Variable Holdoff Trigger inhibit period after end of sweep ad...

Page 7: ...imately 500 kHz Polarity Reversal CH 2 invert Maximum Input Voltage 400 V dc ac peak HORIZONTAL AMPLIFIER Input through channel 1 input X Y mode CH 1 X axis CH 2 Y axis Sensitivity Same as vertical ch...

Page 8: ...splay updated continually Refresh Stored data and display updated by triggered sweep Hold Freezes channel 1 and channel 2 data immediately Save CH 2 Freezes channel 2 data immediately Pretrigger Stora...

Page 9: ...X Y used in conjunction with the X Y control and Trigger SOURCE switch to enable X Y display mode DUAL Displays the channel 1 and channel 2 signals simul taneously Dual trace mode may be either alter...

Page 10: ...el 1 and channel 2 sweeps are chopped and displayed simultaneously normally used at slower sweep speeds When it is pushed in the two sweeps are alternately displayed one after the other normally used...

Page 11: ...s MIN inum normal The holdoff period increases pro gressively with clockwise rotation PULL NORM TRIG When pushed in selects AUTOmatic triggering mode In this mode the oscilloscope generates sweep free...

Page 12: ...GLE switch en gaged this switch readies the scope for single sweep operation A suitabletriggersignalarriving afterpress ing the RESET switch initiates the single sweep 35 READY Indicator Lights when t...

Page 13: ...eform should result EQUIPMENT PROTECTION PRECAUTIONS The following precautions will help avoid damage to the oscilloscope 1 Never allow a small spot of high brilliance to remain stationary on the scre...

Page 14: ...ace should appear on the CRT Adjust the trace brightness with the INTENSITY control and the trace sharpness with the FOCUS control SINGLE TRACE DISPLAY Either channel 1 or channel 2 may be used for si...

Page 15: ...setting up the scope to observe a waveform it provides sweep for wave form observation until other controls can be properly set Once the controls are set operation is often switched back to the norma...

Page 16: ...eep trigger is developed from the trigger source waveform as it crosses the threshold level in a negative going direction TIME BASE Control Set the Time Base TIME DIV control to display the desired nu...

Page 17: ...ms div for complete frames two interlaced fields of video At most points of measurement a composite video signal is of the polarity that is the sync pulses are negative and the video is positive In th...

Page 18: ...ingle sweep mode can be used in the pres ence of very rapidly occurring events even continuous waveforms if desired In that case the period between pushing the RE SET switch and the arrival of the tri...

Page 19: ...ions of waveform 10 24 divisions to be more exact This output takes place at a rate that is 1 10th that of the TIME DIV setting For example if the TIME DIV setting is 10 ms div the plotter output send...

Page 20: ...y fuseholder away from receptacle 3 When reinstalling fuseholder be sure that the fuse is installed so that the correct line voltage is selected see LINE VOLTAGE SELECTION LINE VOLTAGE SELECTION To se...

Page 21: ...ed only as a general check of calibration accuracy not as a signal source for performing recali bration adjustments a voltage standard calibrated at several steps and of 0 3 or better accuracy is requ...

Page 22: ...sweep speed and using X10 magnification This sweep rate is 10 ns div A rise time of less than about five divisions at this sweep speed should be calculated Calculated Measurements For observed rise t...

Page 23: ...quency of the signal at the input jack As an example if three cycles are displayed the calculated frequency would be approximately 15 Hz This is obviously incorrect This occurs because the DSO is taki...

Page 24: ...omers not on open account must include payment in the form of a money order or credit card For the most current flat rate repair charges contact the factory before shipping the product Return all merc...

Page 25: ...t be registered by completing and mailing the enclosed warranty registration card to B K Precision 1031 Segovia Circle Placentia CA 92870 within fifteen 15 days from the date of purchase Exclusions Th...

Page 26: ...n the industry Valuable to those with little knowledge of oscilloscopes as well as the experienced technician or engineer who wishes to refresh their memory or explore new uses for oscilloscopes Downl...

Page 27: ...2000 B K Precision Corp 481 315 9 001 Printed in Taiwan 1031 Segovia Circle Placentia CA 92870...

Reviews: