26
A hexadecimal number between 0 and FFFFh is shown on the current
display.
c)
Use the current coarse and fine knobs to adjust the value displayed by the
power supply until the ammeter indicates the value to 3.0801 A.
d)
Turn the voltage coarse knob to finish the calibration.
While performing the full current calibration, the current display will show
Adc.
After the full current calibration, the message
Done
is shown on the
current display for 1 sec. The power supply does not return to the
Calibration menu because the calibration procedure is finished and the
power supply must be turned off.
6. Leaving the calibration menu
a)
After turning off the power supply, disconnect the power cord.
b)
Remove power supply’s cover.
c)
Remove the calibration jumper.
d)
Reassemble the power supply.
e)
Connect the power cord.
f)
When turned on, the power supply will have the same parameters
programmed and saved before the calibration procedure.
6.4 Instrument Repair Service
Because of the specialized skills and test equipment required for instrument repair
and calibration, many customers prefer to rely upon B&K Precision for this service.
We maintain a network of B&K Precision authorized service agencies for this
purpose. To use this service, even if the instrument is no longer under warranty,
follow the instructions given in the
section of this
manual. There is a nominal charge for instruments out of warranty.