BK Medical Type 8808 User Manual Download Page 9

8808 User Guide

 (BB0339-M)

Imaging with Type 8808

 

9

 

Changing Orientation

To change the orientation of the image on the monitor, refer to the applicable system 
user guide for instructions.

Imaging with Type 8808

The 8808 is designed for simultaneous biplane imaging of the prostate.

Assembling the Transducer for Transrectal Use

The 8808 can be equipped with a water standoff facility to improve contact and field 
of view during transrectal imaging of the prostate. A rubber sheath is mounted on the 
end of the transducer and filled with water to ensure good acoustic contact with the 
rectal wall. By varying the amount of water in the sheath, it is possible to alter the 
position of the transducer’s focal zone in the tissue.

If the water standoff facility is required, it is assembled as illustrated in Figure 4 and 
described as follows:

1

Mount the rubber sheath UA0799 on the transducer and roll it down until it 
covers the groove 

farthest 

away from the tip of the transducer. 

2

Pull O-ring UA0210 over the rubber sheath and position in the groove 

nearest 

the tip of the transducer (Picture 2). The O-ring holds the rubber sheath securely 
in place.

3

Attach a connecting tube with a Luer Lock

TM

 and a 3-way tap to the water 

standoff inlet (Picture 3) and fill the rubber sheath with water using a standard 
disposable syringe (Picture 4).

Air bubbles may appear in the rubber sheath. As air bubbles will disrupt the path of 
the ultrasound beam and cause image artifacts, they must be removed.

Summary of Contents for Type 8808

Page 1: ...For Professional Users Only English BB0339 M June 2012 User Guide Type 8808 Biplane Transducer...

Page 2: ...ucts and services As part of our customer satisfaction program we contact a sample of our customers a few months after they receive their orders If you receive an email message from us asking for your...

Page 3: ...ansducer 8 Changing Frequency 8 Using a Transducer Cover 8 Using the Transducer Control Button 8 Changing Orientation 9 Imaging with Type 8808 9 Assembling the Transducer for Transrectal Use 9 Punctur...

Page 4: ...4...

Page 5: ...Imaging Plane The transducer contains two convex arrays one for transverse T imaging and one for sagittal S imaging see Figure 2 The advantage of being able to view two planes simultaneously with one...

Page 6: ...igure 3 Figure 3 The 8808 and the dummy channel bracket NOTE The dummy channel bracket must be removed before the transducer is prepared for disinfection The bracket can be disinfected using the same...

Page 7: ...nd the exposure time as low as possible WARNING Service and repair of BK Medical electromedical equipment must be carried out only by the manufacturer or its authorized representatives BK Medical rese...

Page 8: ...e cover with 1 to 2 ml of sterile water This improves the screen images by preventing image artifacts caused by air bubbles Pull the transducer cover over the transducer Gel also creates a good acoust...

Page 9: ...is possible to alter the position of the transducer s focal zone in the tissue If the water standoff facility is required it is assembled as illustrated in Figure 4 and described as follows 1 Mount t...

Page 10: ...ansducer will be inserted into the patient without water in the sheath After the transducer is inserted water can be introduced again into the sheath 6 If a biopsy channel bracket UA1257 is being used...

Page 11: ...s dark green UA1257 S14 and UA1257S17E Each version includes needle guide biopsy channel bracket and clamp Full product specifications for the puncture equipment can be found in the Product Data sheet...

Page 12: ...non sterile when supplied and must be disinfected by immersion in a suitable solution and autoclaved The biopsy channel bracket may be damaged during use or processing so it must be checked before use...

Page 13: ...7 Figure 7 Biopsy channel bracket UA1257 mounted on the transducer here shown without condom 5 After you attach the channel bracket pull a sterile transducer cover containing a small amount of steril...

Page 14: ...achment over the tip of the transducer until the lock screw pressure pad meets the steel stud on the side of the transducer The puncture attachment should be correctly positioned see Figure 9 before t...

Page 15: ...one appears Move the transducer until the puncture line transects the target Insert the needle and monitor it as it moves along the puncture line to the target The needle tip echo will be seen as a br...

Page 16: ...erefore you must clean transducers immediately after use Use a suitable brush to make sure that biological material and gel are removed from all channels and grooves See Care Cleaning Safety for clean...

Page 17: ...nsducer is scrapped at the end of its life national rules for the relevant material in each individual land must be followed Within the EU when you discard the transducer you must send it to appropria...

Page 18: ...18 June 2012 8808 User Guide BB0339 M...

Page 19: ......

Page 20: ...LEGAL MANUFACTURER BK Medical ApS Mileparken 34 2730 Herlev Denmark Tel 45 44528100 Fax 45 44528199 Email info bkmed dk...

Reviews: