1.4 How to Set the Brainlab Preset to Default (only on Flex Focus 800 or 

bk5000 (SW 5.14.8))

Whenever available, you should set the “

Brainlab” 

preset as the default imaging preset. This helps 

to give a better overlay in the ultrasound software. Otherwise choose another “

Neuro

” preset as 

default if available.

1. Click on [Image] tab

2. Click on [Advanced] button

3. Click on [Customize] button

4. Navigate down until you find the “Neuro” presets on the left side of the page. Within these 

presets you will find the “Brainlab” preset. 

5. Select it with the cursor, then press [Set Preset as Default] on the right side of the page.

1.5 Where is the Brainlab Preset on SW 5.16.2?

Brainlab Preset has expired on SW 5.16.2. 

For N13C5 (9062) or N11C5s (9063) you should use the Neurosurgery, Gen preset. 
This is default on factory setup. 

For X18L5S (9009), you should use Neurosurgery, Spine preset.  
This is default on factory setup. (This is for USA only)

Summary of Contents for bk5000

Page 1: ...connected to the Brainlab system via the Intraoperative Data port using an Ethernet network cable The communication is performed via the OEM protocol of the BK Medical Ultrasound device IMPORTANT The...

Page 2: ...ransducer Optional bk5000 SW License 5 16 2 Brainlab Interface Includes OEM Neuro N13C5 Transducer N11C5s Transducer X18L5s Transducer Optional Demo keys for End user systems 0000 0000 0000 0026 Neuro...

Page 3: ...k Curve Buzz IGS to the LAN port on the bk5000 Flex Focus 800 2 Start the BK Medical Ultrasound device and Brainlab system bk5000 SW 5 14 8 Flex Focus 800 bk5000 SW 5 16 2 3 Click on Image tab 3 Click...

Page 4: ...window 1 2 Configure OEM Interface bk5000 SW 5 14 8 Flex Focus 800 bk5000 SW 5 16 2 1 Click on OEM SETUP tab 1 Press Ctrl Shift O NOTE On a Flex Focus 800 without real keyboard open the Keyboard tab...

Page 5: ...ation systems Intraoperative Data port and the IP address of the BK Medical Ultrasound device need to be in the same subnet and need to have the same subnet mask You can use e g the same IP address 1...

Page 6: ...ton 3 Click on Customize button 4 Navigate down until you find the Neuro presets on the left side of the page Within these presets you will find the Brainlab preset 5 Select it with the cursor then pr...

Page 7: ...on Advanced Settings tab 5 Open the Inbound Rules tab 6 Check and see if there is an OEM or Console OEM rule that allows port 7915 If there is not you will have to add a new rule 7 Add a new inbound r...

Page 8: ...og 2 Press the Down arrow in the windows dialog to get to the Control Panel 3 Click on Windows Firewall 4 Click on Exceptions tab 5 Check and see if there is an OEM or Console OEM rule that allows por...

Page 9: ...inging between the systems by enabling echo request in ICMP 1 On the Windows Firewall Window go to the Advanced tab 2 Select Settings in the section ICMP 3 Make sure that Allow incoming echo requests...

Page 10: ...e in the Image Size section 4 Checking functionality in Ultrasound Navigation 1 x CONTACT YOUR BRAINLAB AUTHORIZED REPRESENTATIVE AND LET HIM HER TAKE OVER FROM HERE Brainlab Service personnel ONLY 1...

Reviews: