10 of 10

IG36483-5

3 Coexistence with Varian or Master Demo 

license on Flex Focus 800

If you have a Varian License or a Master Demo license on your BK Medical Ultrasound device, there is 
a setting that you need to adjust for every pro package of every transducer:

1. Click on [Image] tab

2. Click on [Advanced] button

3. Click on [Customize]

4. Click on [Split / Size] tab

5. Select e.g. Neuro as Pro Package

6. Mark “Lock Image Size” in the “Image Size” section

4 Checking functionality in Ultrasound 

Navigation 1.x

CONTACT YOUR BRAINLAB AUTHORIZED REPRESENTATIVE AND LET HIM / HER 
TAKE OVER FROM HERE.

Brainlab Service personnel 

ONLY

:

1. Verify that the IP addresses of the BK Medical Ultrasound device and the 

Brainlab system are in the same subnet. Verify that the IP address of the BK 
device is entered in the Settings File of Ultrasound Navigation.

2. Open the Ultrasound Navigation 1.x which is available in the Cranial 

Navigation Workflow 1.2 or higher.

3. Check on the following page whether the Ultrasound device is 

connected correctly (green light).

Summary of Contents for bk5000

Page 1: ...connected to the Brainlab system via the Intraoperative Data port using an Ethernet network cable The communication is performed via the OEM protocol of the BK Medical Ultrasound device IMPORTANT The...

Page 2: ...ransducer Optional bk5000 SW License 5 16 2 Brainlab Interface Includes OEM Neuro N13C5 Transducer N11C5s Transducer X18L5s Transducer Optional Demo keys for End user systems 0000 0000 0000 0026 Neuro...

Page 3: ...k Curve Buzz IGS to the LAN port on the bk5000 Flex Focus 800 2 Start the BK Medical Ultrasound device and Brainlab system bk5000 SW 5 14 8 Flex Focus 800 bk5000 SW 5 16 2 3 Click on Image tab 3 Click...

Page 4: ...window 1 2 Configure OEM Interface bk5000 SW 5 14 8 Flex Focus 800 bk5000 SW 5 16 2 1 Click on OEM SETUP tab 1 Press Ctrl Shift O NOTE On a Flex Focus 800 without real keyboard open the Keyboard tab...

Page 5: ...ation systems Intraoperative Data port and the IP address of the BK Medical Ultrasound device need to be in the same subnet and need to have the same subnet mask You can use e g the same IP address 1...

Page 6: ...ton 3 Click on Customize button 4 Navigate down until you find the Neuro presets on the left side of the page Within these presets you will find the Brainlab preset 5 Select it with the cursor then pr...

Page 7: ...on Advanced Settings tab 5 Open the Inbound Rules tab 6 Check and see if there is an OEM or Console OEM rule that allows port 7915 If there is not you will have to add a new rule 7 Add a new inbound r...

Page 8: ...og 2 Press the Down arrow in the windows dialog to get to the Control Panel 3 Click on Windows Firewall 4 Click on Exceptions tab 5 Check and see if there is an OEM or Console OEM rule that allows por...

Page 9: ...inging between the systems by enabling echo request in ICMP 1 On the Windows Firewall Window go to the Advanced tab 2 Select Settings in the section ICMP 3 Make sure that Allow incoming echo requests...

Page 10: ...e in the Image Size section 4 Checking functionality in Ultrasound Navigation 1 x CONTACT YOUR BRAINLAB AUTHORIZED REPRESENTATIVE AND LET HIM HER TAKE OVER FROM HERE Brainlab Service personnel ONLY 1...

Reviews: