8808e User Guide
(BB1856-F)
Puncture and Biopsy Facilities
11
The puncture line for UA1322-S, UA1322-S14 and UA1326 on 8808e is shown in
Figure 5. UA1322-S and UA1326 have a bore diameter of 1.6 mm, suitable for 17-
gauge needles. UA1322-S14 has a bore diameter of 2.1 mm, suitable for 14-gauge
needles.The puncture line is angled at 19° to the transducer’s axis.
Figure 5. Illustration of the puncture line for biopsy guide UA1322-S, UA1322-S14 and UA1326.
Mounting a Transrectal Biopsy Guide
To mount a biopsy guide on 8808e:
1
If a dummy channel bracket is mounted on the transducer, remove it.
2
Pull a transducer cover or a standard condom containing a suitable amount of
imaging gel over the transducer.
3
Slide the needle guide into the biopsy channel bracket.
4
Insert the assembled needle guide and biopsy channel bracket into the open
channel on the transducer. A small nodule on the end of the channel bracket fits
into an indentation in the channel on the transducer to help you place the bracket
correctly. Click the channel bracket into position on the transducer and lock it
into place (see Figure 6).
Figure 6. Biopsy channel bracket and needle guide mounted on 8808e.
WARNING
Do not use excessive force when inserting the needle guide.
WARNING
Ensure that the channel bracket and needle guide are correctly positioned. Never insert the
needle guide while the transducer is inside the patient.