background image

 

 ML0030   August 2009                                                                                Copyright 2009   Bitronics, LLC 

 

− 

5

 −

 

otherwise occur during the measurement of distorted waveforms with non-true RMS 
instruments.  Voltage of a given phase is sampled first, followed by the current of the same 
phase.  Phases A,B and C are sampled in succession, providing the MCU with 
instantaneous measurements of all voltage and current inputs.  Samples are accumulated 
for 0.5 seconds, at which time the MCU calculates the WATTs and VARs for each phase.  
The VARS quantity for each phase is derived from a power triangle calculation where the 
WATTS and VAs are known.  This technique provides a "true" measure of VARs even with 
distorted waveforms.  Zero offset is also adjusted for each signal channel every 0.5 
seconds by the MCU.  These per phase quantities are then summed to form the total three 
phase WATTS and VARS.  Calibration constants stored in both the Power Supply 
EEPROM and the EEPROM located on this board provide drift-free calibration, and 
complete interchangeability of Analog Processing boards.  Checksums are incorporated 
into both EEPROMs which are read periodically by the microcontroller to check the integrity 
of the calibration constants and the CT and PT setting (See section 4.3 ).  A "Master Gain" 
trimpot is also located on the AP board to provide the user with fine tuning capability if it is 
necessary to match other devices on the power system.  Once the total WATTS and VARS 
have been calculated, the MCU scales the values by the external PT and CT ratios which 
have been selected by the user, and displays the values. 
 
A second function of the AP board is to provide the 0 to +/-1 milliamp transducer output(s).  
This is accomplished with 12-bit digital to analog converters (DAC), one for each output 
channel.  Constants stored in the EEPROM on the analog board are read by the MCU and 
used to compensate for gain and offset of each output channel.  This technique provides 
stable calibrations for the output(s).  Additional circuitry converts the output of each DAC to 
the 0 to +/-1 milliamp current.  This additional circuitry is powered from a +15volt and a 
-15volt supply also located on the AP module.  The outputs have a +/- 10volt compliance, 
and can drive a +/-1.0mA into a 10k ohm load, with a reduced load the output can drive up 
to +/-2.0mA.  An indep5 volt power supply which provides an electrically "quiet" 
supply for all the analog circuitry, and a high precision low drift reference are also present 
on the AP board.  Both the reference and 5volt supply are shared by the ADC and DAC 
circuitry.   
 

2.6  Microcontroller Board (MCU Board)

  

 
The microcontroller board consists of an Intel 80C51FA microcontroller (MCU), address 
latch, EPROM memory and a watchdog timer.  All the data acquisition, signal processing 
and display manipulation are controlled by the microcontroller.  Communications to the 
other boards is accomplished via a serial data link comprising a set of three lines common 
to all the other devices (ADC, 2 DACs, 2 EEPROMs, 2 Display Drivers).  Individual select 
lines for each individual device, allow the MCU to communicate with one device at a time.  
The watchdog timer prevents the MCU from "locking up" in the event of a transient or other 
type of interference.  The watchdog timer also provides a reset on power-up or when 
resuming from a brownout (low supply).  The watchdog timer can be triggered manually, by 
entering the CT/PT set mode (See section 4.1) and holding down the select push button for 
approximately 1.2 seconds.  In the unlikely event of a microcontroller failure, the watchdog 
circuit will continuously attempt to restart the processor.  A positive indication of this 
condition is provided by having the watchdog flash both of the LED displays on the front 
panel.  
 

Summary of Contents for QTWIE2

Page 1: ...Instruments Watt VAR User Manual 2 Element 3 Wire August 7 2009 ML0030 Document Revision B 2009 by Bitronics LLC Includes 0 1mA Transducer Outputs No QTWIE2 WATT VAR METER No WTWIE3 WATTMETER No RTWIE3 VARMETER ...

Page 2: ...rent and Potential Transformer Board CT PT Board 4 2 4 Power Supply Board PS Board 4 2 5 Analog Processing Board AP Board 4 2 6 Microcontroller Board MCU Board 5 2 7 LED Display Board LED Board 6 3 0 INSTALLATION 7 3 1 Initial Inspection 7 3 2 Power Requirements 7 3 3 Overcurrent Protection 7 3 4 Mains Disconnect 7 3 5 Instrument Mounting 7 3 6 Surge Protection 8 4 0 FIELD ADJUSTMENTS 9 4 1 Rescal...

Page 3: ...d by Bitronics the protection provided by the equipment may be impaired WARRANTY AND ASSISTANCE Products manufactured by Bitronics LLC are warranted against defects in materials and workmanship for a period of thirty six 36 months from the date of their original shipment from the factory Products repaired at the factory are likewise warranted for eighteen 18 months from the date the repaired produ...

Page 4: ......

Page 5: ...of input signal levels WATT VAR meters are modular in design with push button rescaling to display primary values when using any standard current and voltage transformer Rescaling can be done in the field in a matter of minutes without removing the instrument from the panel or the need for any calibration equipment Physical dimensions are the standard 4 round case with an overall length of 6 5 inc...

Page 6: ...d by CT PT ratio Scaling User selectable using internal CT PT tables Accuracy 0 25 Class ANSI Std 460 1988 Signal Frequency 50Hz to 400Hz Including Harmonics Current Output 0 to 1mA into 10K ohms or less 0 4 ripple p p or less Calibrated at 1000W at meter input Overload to 2mA into 5K ohms or less Power Requirements 115 Vac 20 6 VA 230 Vac or Universal AC DC optional Fuse 1 5 Ampere non time delay...

Page 7: ...he need to remove the meter from the panel or to remove the meter from service see section 4 4 Detailed descriptions of each of the boards can be found in the following sections 2 2 Input Signal Connections The WATT VAR Meters have six independent signal inputs one current and one voltage for each phase being measured Current and voltage signals are connected directly to 10 32 brass studs on the r...

Page 8: ...500Vac between the inputs and any output 2 4 Power Supply Board PS Board The power supply circuit is a conservative conventional design Low drop rectifiers and a low drop out solid state regulator minimize internal power dissipation Filter capacitors are operated at a fraction of their voltage and temperature ratings and should provide years of trouble free service under extreme environmental cond...

Page 9: ...stored in the EEPROM on the analog board are read by the MCU and used to compensate for gain and offset of each output channel This technique provides stable calibrations for the output s Additional circuitry converts the output of each DAC to the 0 to 1 milliamp current This additional circuitry is powered from a 15volt and a 15volt supply also located on the AP module The outputs have a 10volt c...

Page 10: ...en segment decoder driver chip which accept serial data from the MCU and decodes the data into the seven segment and digit select outputs necessary for the multiplexed display The high current cathode drive is provided by an MC1472 driver for each pair of digits On power up or any other time the MCU is reset a display test will be conducted that displays 1 8 8 8 on the top display followed by 1 8 ...

Page 11: ...nstrument The power supply and regulators provide constant dc power to the modules independent of variations in ac supply voltage over this range If the supply voltage drops below the point at which the regulators can function properly the watchdog timer will cause the displays to flash as described previously 3 3 Overcurrent Protection A UL listed 1 5 Ampere non time delay M fuse is to be series ...

Page 12: ...s of the instrument The GREEN lead of the MOV assembly should be connected to a good earth ground In most instances this is usually accomplished by connecting the GREEN lead to the indicated front mounting stud This mounting stud is a safety ground for the instrument and should be connected to a protective earth circuit refer to Figure 3 Although the Line to Ground capacitors are 3kV and UL rated ...

Page 13: ... CT or PT will be shown on the top display All CT ratios are assumed to have 5 amps as their secondary maximum outputs Therefore 10 0 on the display corresponds to a 10 5 CT PTs are represented by a ratio to 1 so 20 0 on the display would correspond to a 20 1 PT Powers of ten can also be set as the decimal point moves from left to right by using the select button 3 Index through the available CT P...

Page 14: ... relay ratio 1 20 12 0 120 1 40 14 0 140 1 50 15 0 150 1 73 17 3 173 relay ratio 1 75 17 5 175 1 80 18 0 180 2 00 20 0 200 2 31 23 1 231 relay ratio 2 40 24 0 240 2 50 25 0 250 2 51 25 1 251 relay ratio 3 00 30 0 300 3 46 34 6 346 relay ratio 3 50 35 0 350 4 00 40 0 400 4 35 43 5 435 4 50 45 0 450 5 00 50 0 500 5 77 57 7 577 relay ratio 6 00 60 0 600 6 35 63 5 635 6 64 66 4 664 6 93 69 3 693 relay...

Page 15: ...o decrease the indicated measurement NOTE Because these WATT VAR meters are a full 2 elements the instrument can be calibrated with a single phase source all current inputs should be connected in series and all potential inputs should be connected in parallel 5 Remove the screwdriver and replace the calibration seal 6 Replace gasket front cover and four cover screws The 0 to 1mA output will track ...

Page 16: ...e CT and or PT ratios The transducer outputs are still functional and accurate Calibration constants for the CT PT Board are in error The display and the transducer output are reduced in accuracy to approximately 3 Calibration constants for the Analog Processing Board are in error The display and the transducer output are reduced in accuracy to approximately 3 The watchdog timer is attempting to r...

Page 17: ...input quantity exceeds the range of the instrument Both display and transducer output accuracy reduced by an amount depending upon the degree of over range The microcontroller has detected a fault in program memory and is attempting to restart itself Microcontroller will force transducer outputs to 2ma 0 1ma out or 2ma 4 20ma out Instrument cannot read any signals Microcontroller will force transd...

Page 18: ...cated at the top of the module pull gently on the bail and the Electronics Module will pull out a slight rocking motion may be required Be careful not to bend any of the connector pins in the Base Module CAUTION when the Electronics Module is removed with the instrument powered instrument power 115Vac or 230Vac is present on the circuit boards that remain in the Base Module DO NOT touch or insert ...

Page 19: ...4 HI and LO are marked on the inputs Does polarity matter Yes Correct wiring with proper polarity is essential for proper operation 5 Can I put WATT VAR meters in an outdoor cabinet Yes Many Bitronics meters are installed that way The temperature range of 20C to 70C covers most applications The case is gasketed but not waterproof so it must be placed within an enclosure that provides ingress prote...

Page 20: ...Figure 4 Typical connection diagram for 2 Element Watt Var Watt Var Meters ...

Page 21: ...1 Brodhead Road Bethlehem PA 18017 610 997 5100 Fax 610 997 5450 www novatechweb com bitronics Revision Date Changes By A 01 30 2009 Update Bitronics Name Logo E Demicco B 08 14 09 Updated logos and cover page MarCom C ...

Reviews: