background image

BT-36 Operation manual                                                              11 

P/N:36-ENG-OPM-EXP-R00 

Bistos 

2022.03 

3.5.

 

Object temperature measurement 

 

Turn on the device with pressing START button 

 

Check the measurement mode whether is object  temperature measurement 

mode(

).   

 

Place the device 2 ~ 5 cm distance from the object that you want to measure 
temperature. 

 

Start measurement by pressing START button(

). 

When the measurement is complete, you can hear beep sound and the measured 
value is displayed on screen. If the device sound is off, wait for a second after you 
press the button. If the measured value within the measurable temperature range, 
then the blue light is illuminated with the measured value displayed. If the 
measured value exceeds the measurable temperature range, then the red light is 
illuminated with the error(Err). 
(When measuring, any movement may affect the temperature reading.) 

 

3.6.

 

Memory mode 

 

Press the 

  Memory mode button to switch to memory mode.   

 

The temperature is stored in memory from No. 0 to No. 9 in order from the latest 
data to the previous data. 

 

If there is no other input for a certain period of time, it goes back to Ready screen. 

 

Results of object temperature and errors are not stored in memory. 
 

3.7.

 

Power off 

The power is automatically shut down if there is no other input about 15 seconds after 
measurement.

 

 

3.8.

 

Battery replacement 

When lack of the battery icon(

) displayed on right top of the screen, the battery 

should be replaced. (It may be inaccurate if the user measures the temperature value.)

 

 
 
 
 
 
 

Summary of Contents for BT-36

Page 1: ...INFRARED THERMOMETER BT 36 Operation Manual In order to properly use this product please be sure to read the user s manual before use P N 36 ENG OPM EXP R00 This product is a medical device...

Page 2: ...The contents described in manual does not provide any additional warranties Bistos Co Ltd is not responsible for technical errors editing errors or omissions in the contents contained in manual The in...

Page 3: ...aration before use 9 3 2 Power on 10 3 3 Mode selection 10 3 3 1 Body Object temperature measurement mode 10 3 3 2 Sound On Off 10 3 3 3 Setting 10 3 4 Body temperature measurement 10 3 5 Object tempe...

Page 4: ...by rewarding from measured skin temperature in body temperature measurement mode The recommended part of BT 36 body temperature measurement mode is from 2 5 cm by the center of forehead BT 36 operate...

Page 5: ...gainst the ingress of solid object and liquid IP2X is protection against solid foreign object like a finger IPX2 is protection against some falling water drops vertically when enclosure tilted up to 1...

Page 6: ...and Storage for cleaning and disinfection Be cautious not to get dust or especially metallic objects into the device Do not place the device where it is too humid or changes temperature drastically Do...

Page 7: ...ates to the thermometer it can cause malfunction and damage When measuring temperature check the temperature measurement mode During measurement if you look at the LCD screen while pressing the button...

Page 8: ...is icon will show up on the screen and goes to Sound Off status If you press again it goes to Sound On status Default Sound On MODE selection button This button is to change the measurement mode If th...

Page 9: ...emperature or object temperature without any contacts is placed in this part Battery cover This device uses 2 AAA type batteries Press the top of the cover on the back of the device to open it insert...

Page 10: ...rature range 1 Human body 34 43 93 2 109 4 2 Object 15 50 59 122 Low battery icon If the battery voltage is low a Low battery icon is displayed on the upper right of the LCD screen to indicate that th...

Page 11: ...erature measurement Turn on the device with pressing START button Check the measurement mode whether is body temperature measurement mode Place the device 2 5 cm distance from center of forehead Start...

Page 12: ...measured value displayed If the measured value exceeds the measurable temperature range then the red light is illuminated with the error Err When measuring any movement may affect the temperature read...

Page 13: ...y place for at least 12 hours If the measurement is not working properly When measuring infants or children if they struggle too much measurement errors may occur because the distance between the skin...

Page 14: ...ment mode Measuring method Infrared sensing of skin surface Non contact measurement Display method Mono LCD Expected lifetime 1 year Power Battery 3V 1 5V AAA battery 2EA Standard configuration User M...

Page 15: ...ssions or decreased electromagnetic immunity of this device and result improper operation Electric medical devices require special precautions regarding EMC and should be used in accordance with the E...

Page 16: ...4 2 8 kV Contact 2 kV 4 kV 8 kV 15 kV Air 8 kV HCP 8 kV VCP Complies Floors should be wood concrete or ceramic tile If floors are covered with synthetic material the relative humidity should be at le...

Page 17: ...ered by batteries only Surge IEC 61000 4 5 1 kV line s to line s 2 kV line s to earth Not applicable Power frequency 50 Hz and 60 Hz magnetic field IEC 61000 4 8 30 A m Complies Power frequency magnet...

Page 18: ...ess Contact Seller Bistos Co Ltd Manufacturer Bistos Co Ltd Thank you for purchasing the Contactless infrared thermometer This product has passed through strict quality control and inspection The comp...

Reviews: