Maintenance
21
21. Adjust R6 until the 4304A meter indicates 400.
Figure
11
Range Adjustment Potentiometers
Storage
When storing the Bird 4304A, turn the element so that the arrow points down
and set the range switch to 5 W. This will shunt the meter circuit and protect the
meter by dampening needle action.
Customer Service
Any maintenance or service procedure beyond the scope of those in this
chapter should be referred to a qualified service center.
If the unit needs to be returned for any reason, request an Return Material
Authorization (RMA) through the Bird Technologies website. All instruments
returned must be shipped prepaid and to the attention of the RMA number.