BioVision Technologies CMAR-01 Quick Start Manual Download Page 3

3

BVT93569-00001-B

The BioVision CMAR-01 is sterilizable by low temperature methods. 

The only recommended sterilization method is Ethylene Oxide (EtO). EtO 
is done at low temperatures and has excellent Sterility Assurance Levels 
(SAL).

Using a high temperature sterilization method such as autoclave or steam; 
will cause irreparable damage and is not covered by the warranty. 

The camera head, cables, and connectors are not immersible.  Sterilizing 
and disinfecting soaks are very corrosive and may cause damage to the 
camera system. This damage is not covered by the warranty. 

To ensure your safety and the safety of others around you, always follow 
the manufacturer’s instructions when using a sterilizer.

To ensure the optical performance of your 
camera system, clean the optical areas 
indicated by arrows in the illustration to the 
right.

The recommended method is a cotton 
swab or gauze damp with alcohol.  Lightly 
rub these areas until any foreign material 
is cleared away.

Note that you should not clean inside the 
outer ring of the SurgView

TM

 connector.  

This area has gold-pins and should never 
need cleaning.

Hand clean the rest of the camera with a 
damp cloth and a mild detergent if 
necessary. Do not use any solvents.

DO NOT SOAK

DO NOT AUTOCLAVE