Chapter 5
Programmable Parameters
Evia HF / HF-T Technical Manual
PAGE 60
HF with normal
sinus function
HF with
permanent AF
HF with low
CLS response
HF with medium
CLS response
HF with high
CLS response
Mode
DDD
VVI-CLS
DDD-CLS
DDD-CLS
DDD-CLS
AV Delay 1
150
--
150
150
150
AV Delay 2
140
--
140
140
140
AV Delay 3
130
--
130
130
130
AV Delay 4
120
--
120
120
120
AV Delay 5
110
--
110
110
110
Sense Compensation
-45
--
-45
-45
-45
RA Capture Control
ON
--
ON
ON
ON
Threshold test start
3.0
--
3.0
3.0
3.0
Safety Margin
1.0
--
1.0
1.0
1.0
Search type
Time of Day
--
Time of Day
Time of Day
Time of Day
Time of day
2:00
--
2:00
2:00
2:00
RV Capture Control
ON
ON
ON
ON
ON
Threshold test start
3.0
3.0
3.0
3.0
3.0
Safety Margin
1.0
1.0
1.0
1.0
1.0
Search type
Time of Day
Time of Day
Time of Day
Time of Day
Time of Day
Time of day
2:00
2:00
2:00
2:00
2:00
LV Capture Control
ON
ON
ON
ON
ON
Threshold test start
3.0
3.0
3.0
3.0
3.0
Safety Margin
1.0
1.0
1.0
1.0
1.0
Search type
Time of Day
Time of Day
Time of Day
Time of Day
Time of Day
Time of day
2:00
2:00
2:00
2:00
2:00
PMT
ON
--
ON
ON
ON
High Atrial Rate
Mode Switch
--
Mode Switch
Mode Switch
Mode Switch
High Ventricular Rate
ON
ON
ON
ON
ON
Patient trigger
OFF
OFF
OFF
OFF
OFF
Pre Trigger Recording
75
75
75
75
75
IEGM signal
Filtered
Filtered
Filtered
Filtered
Filtered
HAR limit
200
--
200
200
200
HVR limit
160
160
160
160
160
HVR counter
8
8
8
8
8
Thoracic Impedance
ON
ON
ON
ON
ON
Home Monitoring
ON
ON
ON
ON
ON
Time of Transmission
AUTO
AUTO
AUTO
AUTO
AUTO
Table 17: ProgramConsult
5.18 Home Monitoring (Evia HF-T)
Home Monitoring enables the exchange of information about a patient’s cardiac status from the implant
to the physician. Home Monitoring can be used to provide the physician with advance reports from the
implant and can process them into graphical and tabular format called a Cardio Report. This information
helps the physician optimize the therapy process, as it allows the patient to be scheduled for additional
clinical appointments between regular follow-up visits if necessary.
Summary of Contents for Evia HF
Page 1: ...Cardiac Rhythm Management Heart Failure Therapy Evia HF HF T Evia HF HF T Technical Manual...
Page 8: ...Chapter Table of Contents Evia HF HF T Technical Manual PAGE vi...
Page 12: ...Chapter 2 Indications Evia HF HF T Technical Manual PAGE 4...
Page 14: ...Chapter 3 Contraindications Evia HF HF T Technical Manual PAGE 6...
Page 98: ...Chapter 7 Product Storage and Handling Evia HF HF T Technical Manual PAGE 90...
Page 104: ...Chapter 9 Lead Connection Evia HF HF T Technical Manual PAGE 96...
Page 118: ...Chapter 10 Elective Replacement Indication ERI Evia HF HF T Technical Manual PAGE 110...
Page 130: ...Chapter 13 Order Information Evia HF HF T Technical Manual PAGE 122...
Page 132: ...Chapter 14 Order Information Evia HF HF T Technical Manual PAGE 124...
Page 136: ...PAGE 128 Appendix A Evia HF HF T Technical Manual...