2. Detection and Configuration of the Lead Polarity
The Evia CRT-P switches the lead polarity to bipolar immediately after a bipolar lead is detected.
When the lead impedance is between 100 and 2500 Ohms, the lead connected is classified as bipolar
and the sense and pace polarities are set appropriately. The device switches back to unipolar if the
lead impedance falls outside this range.
3. 10-Minute Confirmation Phase
A 10-minute confirmation phase is initiated after detection of the lead. The lead impedance
measurement is performed alternating between the atrial and ventricular channel. Additional
impedance measurements at the end of this phase confirm the lead detection and lead polarity. The
impedance measurements need to fall within the range of 100 to 2500 Ohms.
A lead impedance measurement outside this range restarts the confirmation process.
Interrogation of the pulse generator during the confirmation phase results in a message that implant
detection is activated. The confirmation phase is terminated and the pulse generator features are
activated if it is reprogrammed during this time.
4. Activation of Pulse Generator Features
The following pulse generator features are activated after the confirmation phase has been
successfully completed:
• Auto Lead Check
• Capture Control
• Statistics
• Rate response
• Collection of patient-specific impedance waveform characteristics for adapting the CLS
algorithm to the patient (data does not control the pacing rate until CLS is programmed on)
• PMT Management
• Auto PVARP
• 2:1 Lock-In protection
Summary of Contents for Evia HF
Page 1: ...Cardiac Rhythm Management Heart Failure Therapy Evia HF HF T Evia HF HF T Technical Manual...
Page 8: ...Chapter Table of Contents Evia HF HF T Technical Manual PAGE vi...
Page 12: ...Chapter 2 Indications Evia HF HF T Technical Manual PAGE 4...
Page 14: ...Chapter 3 Contraindications Evia HF HF T Technical Manual PAGE 6...
Page 98: ...Chapter 7 Product Storage and Handling Evia HF HF T Technical Manual PAGE 90...
Page 104: ...Chapter 9 Lead Connection Evia HF HF T Technical Manual PAGE 96...
Page 118: ...Chapter 10 Elective Replacement Indication ERI Evia HF HF T Technical Manual PAGE 110...
Page 130: ...Chapter 13 Order Information Evia HF HF T Technical Manual PAGE 122...
Page 132: ...Chapter 14 Order Information Evia HF HF T Technical Manual PAGE 124...
Page 136: ...PAGE 128 Appendix A Evia HF HF T Technical Manual...