the system at the patient’s home may
require installation of adapters and/or
analog phone lines. Please contact your
telephone service provider for informa-
tion on the type of phone line that you
have.
Light 4
Off –
When this light is off
even though
the
CardioMessenger is in the charging station,
proceed as follows:
Solutions –
Check if light 5 on the charging
station is illuminated. Is this not the case,
refer to the solutions described for light 5.
If light 5 is illuminated, check whether the
connecting parts on the CardioMessenger or
on the charging station are easily accessible
and not dirty.
If light 4 is still off, inform your physician
so that the device can be checked and the
battery may be exchanged.
Note:
If you take the CardioMessenger
out of the charging station, the battery
monitor light will go out to save energy.
Light 5
Off –
If this light is off, proceed as follows:
Solutions –
Check that the plugs of the
power cables are connected both to the
outlet in the wall and to the power supply.
Check that the green light of the power
supply is illuminated. If this is not the case,
either the power outlet is not being supplied
with electricity or the power supply is defec-
tive. Return a non-functioning power supply
to your physician.