Safety during Use
Electromagnetic Interferences
7
Section
Test
Test level
8.9
IEC 61000‑4‑4
Transient conducted surge voltages (EFT,
bursts)
•
± 2 kV mains supply
•
± 1 kV signal line
IEC 61000‑4‑5
Surge voltage waves on supply lines
•
± 0.5 kV, ± 1 kV line to line
IEC 61000‑4‑6
Conducted radiofrequency interference
•
3 V
•
6 V in ISM bands
IEC 61000‑4‑8
AC frequency magnetic fields
•
30 A/m
•
50 Hz
IEC 61000‑4‑11
Voltage fluctuations and interruptions in
supply voltage
•
0% U
T
; 0.5 cycles
At 0°, 45°, 90°, 135°, 180°, 225°, 270°,
315°
•
0% U
T
; 1 cycle
At 0°
•
70% U
T
; 25 cycles, 30 cycles
At 0°
•
0% U
T
; 250 cycles, 300 cycles
Strong electromagnetic interferences that occur in the close vicinity of electrical motors, power cables,
PCs, monitors or other – possibly defective – electrical devices may compromise the function of the
device in certain cases.
This kind of device malfunction should be considered as a possible cause if the following is observed:
•
The device switches on by itself.
•
The device displays other inexplicable behaviors.
Correct operation of the device can be restored by the following methods:
•
Switch off the interfering electrical device.
•
Remove the source of interference from this device.
•
Switch this device off and on.
If the interference continues, contact BIOTRONIK immediately.
WARNING
Risk of electromagnetic interference through the use of unauthorized accessories
The use of accessories, transducers, or cables not listed by BIOTRONIK or of accessories other
than those specified by BIOTRONIK, can produce elevated electromagnetic emissions or cause
degradation in the device's resistance to electromagnetic interference. Such effects can lead to the
faulty operation of the device.
•
Only use accessories authorized by BIOTRONIK.