Safety during Use
Electromagnetic Interferences
6
Electromagnetic Interferences
Possible electromagnetic interference
This device is protected from electromagnetic interference and electrostatic discharges in a medical
practice. At the same time, the emitted interference is reduced to a minimum. The device thus meets
all requirements of IEC 60601‑1‑2.
The following tests were performed according to IEC 60601-1-2: 2014:
Section
Test
Test level
7.1
EN 55011 (CISPR 11)
Conducted interference emissions
•
Group 1
•
Class B
EN 55011 (CISPR 11)
Radiated emission
7.2.1
IEC 61000‑3‑2
Harmonic distortion (harmonic currents
in the mains supply)
•
Class A
7.2.2
IEC 61000‑3‑3
Voltage fluctuations and flicker in the
mains supply
8.9
IEC 61000‑4‑2
Electrostatic discharge (ESD)
•
± 8 kV contact discharge
•
±15 kV air discharge
8.9/8.10
IEC 61000‑4‑3
Electromagnetic fields
•
Modulation: 1 kHz
•
3 V/m, 80 MHz – 2.7 GHz
•
Limits for RF communication
equipment per Table 9 in
IEC 60601-1-2 (9 – 28 V/m)