4. Select a prime Rate in µL/secon.
5. Click
Purge
to start the process.
Plate Shaking Options
The Synergy H1 supports multiple plate shaking options, as described below. Shaking is
controlled using Gen5 by adding a Shake step to a protocol's procedure.
Figure 3-3: Gen5 Shake Step options
Mode
Speed
Amplitude
(in 1-mm steps)
Frequency
Linear
-
1 mm to 6 mm
18 Hz to 6 Hz
Orbital
Slow
1 mm to 6 mm
10 Hz to 3 Hz
Orbital
Fast
1 mm to 6 mm
14 Hz to 5 Hz
Double Orbital
Slow
1 mm to 6 mm
~10 Hz to ~3 Hz
Double Orbital
Fast
1 mm to 6 mm
~14 Hz to ~5 Hz
Note: Frequency is based on the Amplitude selected
38 | Chapter 3: Getting Started
BioTek Instruments, Inc.