Biomagnetic Separation - Magnetic Bead Assays | 145
Biomagnetic Separation - Magnetic Bead Assays
The microplate carrier, when the magnet adapter is used, supports placement of a
magnet under the microplate. The magnet induces magnetic beads to settle at the
bottom of the wells to help retain them during a wash protocol’s aspirate cycle.
The MultiFlo FX supports standard microplates and these magnets available for
purchase from BioTek:
Plate Type Magnet
PN
96-well
96 Flat Magnet
7103016
96 Ring Magnet
7102216
384-well
384 Flat Magnet
7103017
384 Ring Magnet 7102215
You can use a different magnet if it fits in the carrier and accommodates your
plates. Contact BioTek TAC or visit the Customer Resource Center at
www.biotek.com to obtain a drawing of the carrier with its dimensions.
Persons with pacemakers/implants should avoid direct contact.
Keep all
magnetic media, watches, and sensitive electronic devices away from the magnets.
Credit cards, tapes, and disks can be erased in the presence of a magnetic field. Bodily
harm [pinching of hands and skin] can result if magnets are not handled correctly.
Maintain distance between two or more magnets.
Take advantage of the sample magnetic-bead protocols shipped onboard the MultiFlo
FX. Copy W-Luminex_Mag_Flat_96, for example, and modify the parameters to suit your
assay requirements.
Handling and Cleaning the Magnets
For best magnet strength and bead retention, the bottom of the microplate must be
as close to the magnet as possible. We recommend using flat-bottom plates with
minimal support “webbing” between the sides of the outer wells and the plate
skirts.
Handle the magnets with care. Avoid direct contact with the magnet material. Keep
loose ferrous material away and do not attempt to disassemble.
BioTek Instruments, Inc.