3: Install the MicroFlo’s Components
|
15
MicroFlo Select Operator's Manual
Tubing Cassette Diagram
Figure 4: Cassette Diagram
Important: remove the tip guard from the Tip Holder before installing
the dispense cassette.
1.
Tip Holder
: The cassette’s easiest part to identify, the tip holder fits into the
dispense arm to the right of the pump for positioning above the plate. The 1536
Cassette’s cover plate is made of steel, instead of Ultem® (polyetherimide) like
the other cassette types. Remove the temporary
Tip Guard
before installation.
2.
Center Holder
: The center holder is labeled to identify the size of the cassette
tubing. It also has a serial number for tracking purposes. It fits in between the
tip holder and the tube tensioner and fixes the tubes in place. On the MicroFlo
it slides into grooves on the right side of the pump.
3.
Tube Tensioner
: The transparent 5-mm scale on its front surface identifies the
tube tensioner. It has 8 internal screws for stretching the tubing, one for each
tube. The tube tensioner’s scale is useful when calibrating the cassette.
0.
4.
Tube Organizer
: At the opposite end of the cassette from the tip holder, the
tube organizer holds the 8 tubes together for inserting into the fluid vessel.
1
2
3
4
Tip Guard
(for shipping)
Tip Holder
Center Holder
Tube Tensioner
Tube Organizer
Summary of Contents for MicroFlo select
Page 1: ...Operator s Manual Dispenser MicroFlo Select ...
Page 2: ......
Page 44: ...26 Installation BioTek Instruments Inc ...
Page 74: ...56 Operation BioTek Instruments Inc ...
Page 116: ...98 Dispensing to Test Tubes BioTek Instruments Inc ...
Page 126: ...108 Gravimetric Calibration Instructions MicroFlo Select Microplate Dispenser ...