102
| Chapter 5: Preventive Maintenance
Recommended Maintenance Schedule
A general Preventive Maintenance (PM) regimen for all ELx808 models includes
periodically cleaning all exposed surfaces, inspecting/cleaning the wavelength filters,
replacing the lamp, and decontamination. Robotic models should have certain components
lubricated. The table below recommends Preventive Maintenance tasks and the frequency
with which each task should be performed.
The risk factors associated with your tests may require that some or
all of the procedures be performed more or less frequently than shown
below.
Task
Frequency
Every
three
months
Every six
months
As needed
Before
storage or
shipment
Clean Exposed Surfaces, p. 103
9
Inspect/Clean Wavelength
Filters, p. 104
9
(Optional) Lubricate Robotic
Components, p. 105
9
Replace Lamp and Clean
Contacts, p. 107
after ~500
hours
Decontamination, p. 111
9
Warnings and Precautions
Please read the following before performing any maintenance procedures.
Warning! Internal Voltage
.
Turn off and unplug the
instrument for all maintenance and repair operations.
Warning!
Wear prophylactic gloves when handling
contaminated instruments. Gloved hands should be
considered contaminated at all times; keep gloved hands
away from eyes, mouth, nose, and ears. Eating and drinking
while decontaminating instruments is not advised.
BioTek Instruments, Inc.
Summary of Contents for ELx808
Page 1: ...Operator s Manual Absorbance Microplate Reader ELx808 ...
Page 2: ......
Page 30: ...8 Chapter 1 Introduction BioTek Instruments Inc ...
Page 100: ...78 Chapter 3 Operation BioTek Instruments Inc ...
Page 136: ...114 Chapter 5 Preventive Maintenance BioTek Instruments Inc ...
Page 144: ...122 Chapter 6 Error Codes BioTek Instruments Inc ...
Page 146: ...124 Appendix A Sample Reports Samples with Calls on Matrix Report BioTek Instruments Inc ...
Page 147: ...Curve Fit Report 125 Curve Fit Report ELx808 Operator s Manual ...
Page 148: ...126 Appendix A Sample Reports BioTek Instruments Inc Samples with Calls on Column Report ...
Page 149: ...Column Report without Samples 127 Column Report without Samples ELx808 Operator s Manual ...
Page 150: ...128 Appendix A Sample Reports Panel Report BioTek Instruments Inc ...
Page 151: ...Assay Detail Report 129 Assay Detail Report ELx808 Operator s Manual ...
Page 172: ...150 Appendix C Adjusting the Power Input Voltage Setting BioTek Instruments Inc ...