Replace the Lamp and Clean the Contacts
|
107
Robotic Motor Adjustment Procedure
After lubrication/reassembly, the robotic door needs to be set to the open and closed
positions. Press the following keys to adjust the door.
At the
Main Menu
:
1.
Press the
UTIL
soft key.
2.
Press the
SETUP
soft key.
3.
Press the hidden key between the
Main Menu
and
Previous Screen
keys.
You should see the
Door Adjustment
screen.
4.
Press the
CLOSE
soft key.
5.
Press the
UP
soft key until you just see the door move up. You may have to
press the key many times before the door moves. When the door does
move, press the
DOWN
soft key 4 times. This positions the door lift ram
just behind the door roller in the closed position.
6.
Press the
OPEN
soft key.
7.
Press the
UP
soft key to raise the door to the desired height. Press the
DOWN
soft key if you want to adjust the door downward.
Important!
Do not open the door to its
maximum open position.
8.
Press the
Main Menu
key.
9.
Press the
CLOSE
soft key. This will close the door. The door should rest
closed on the top cover.
Replace the Lamp and Clean the Contacts
There are two types of lamp assemblies. Steps 7 and 8 only pertain to
the newer assembly. Review the photos below to determine which
instructions apply to your reader.
The reader’s lamp, on average, should provide 500 hours of total service before it needs to
be replaced. This equates to approximately six months to one year of use under routine
conditions. The intensity of the lamp will slowly drop over time until the reader’s run-time
self-check detects a low signal and warns the user with a displayed error code. The bulb
should be replaced at this time (PN 3400508), and the electrical contacts in the lamp socket
cleaned.
ELx808 Operator’s Manual
Summary of Contents for ELx808
Page 1: ...Operator s Manual Absorbance Microplate Reader ELx808 ...
Page 2: ......
Page 30: ...8 Chapter 1 Introduction BioTek Instruments Inc ...
Page 100: ...78 Chapter 3 Operation BioTek Instruments Inc ...
Page 136: ...114 Chapter 5 Preventive Maintenance BioTek Instruments Inc ...
Page 144: ...122 Chapter 6 Error Codes BioTek Instruments Inc ...
Page 146: ...124 Appendix A Sample Reports Samples with Calls on Matrix Report BioTek Instruments Inc ...
Page 147: ...Curve Fit Report 125 Curve Fit Report ELx808 Operator s Manual ...
Page 148: ...126 Appendix A Sample Reports BioTek Instruments Inc Samples with Calls on Column Report ...
Page 149: ...Column Report without Samples 127 Column Report without Samples ELx808 Operator s Manual ...
Page 150: ...128 Appendix A Sample Reports Panel Report BioTek Instruments Inc ...
Page 151: ...Assay Detail Report 129 Assay Detail Report ELx808 Operator s Manual ...
Page 172: ...150 Appendix C Adjusting the Power Input Voltage Setting BioTek Instruments Inc ...