(Optional) Lubricate Robotic Components
|
105
7.
Clean the filters using lens-cleaning tissue moistened with a small amount of
isopropyl, ethyl, or methyl alcohol. Ensure that the filters remain in their
current locations.
8.
Replace the filter wheel on its magnetic hub. Ensure that the peg on the wheel
lines up with the notch on the hub, and that the filter wheel is positioned flat
against the hub and rotates freely.
9.
Close the top shroud and replace the seven perimeter screws.
Store filters in a dry environment, such as a desiccator, if they will not
be used for an extended period.
If the reader continues to fail tests in
Chapter 4
,
Instrument
Qualification
, after you have cleaned the filters and the lamp
contacts, or if one or more filters appear to have deteriorated, contact
your BioTek Representative or BioTek.
(Optional) Lubricate Robotic Components
This section applies to ELx808 robotic models only. You will need lubricant, BioTek
PN 66039.
1.
Remove the top shroud of the instrument (see instructions in
Chapter 2
).
2.
Lubricate the components as shown in
and
every six
months, or after 10,000 cycles, according to the instructions in the table below.
Component
Location
Procedure
A
Lubricate the underside of the bracket.
B
Lightly lubricate the shaft indicated.
C
Apply a heavy coating of lubricant to both sides of the motor
shaft. Run the motor shaft back and forth through the motor to
ensure that the internal drive nut is heavily lubricated. Use the
robotic motor adjustment procedure described on the next
page to work the lubricant into the internal drive nut.
D
Lightly lubricate the surface of the roller.
E
Lightly lubricate the underside of the hook-in bracket.
3.
Reattach the top shroud.
ELx808 Operator’s Manual
Summary of Contents for ELx808
Page 1: ...Operator s Manual Absorbance Microplate Reader ELx808 ...
Page 2: ......
Page 30: ...8 Chapter 1 Introduction BioTek Instruments Inc ...
Page 100: ...78 Chapter 3 Operation BioTek Instruments Inc ...
Page 136: ...114 Chapter 5 Preventive Maintenance BioTek Instruments Inc ...
Page 144: ...122 Chapter 6 Error Codes BioTek Instruments Inc ...
Page 146: ...124 Appendix A Sample Reports Samples with Calls on Matrix Report BioTek Instruments Inc ...
Page 147: ...Curve Fit Report 125 Curve Fit Report ELx808 Operator s Manual ...
Page 148: ...126 Appendix A Sample Reports BioTek Instruments Inc Samples with Calls on Column Report ...
Page 149: ...Column Report without Samples 127 Column Report without Samples ELx808 Operator s Manual ...
Page 150: ...128 Appendix A Sample Reports Panel Report BioTek Instruments Inc ...
Page 151: ...Assay Detail Report 129 Assay Detail Report ELx808 Operator s Manual ...
Page 172: ...150 Appendix C Adjusting the Power Input Voltage Setting BioTek Instruments Inc ...