Biotage ELSD-1080
3.
Photodetector window
f
Displays current detector output
f
Displays current gain setting
f
Auto-zero the detector
Figure 6. Control Software Main Screen
3.5.2
Control Software-Parameter Settings
All of the six basic detector control parameters can be modified
via the detector control panel. Changes to temperature and gas
flow are made via the METHOD button, while changes to the
LED, Smoothing and Gain are adjusted via the Edit menu: -
(EDIT
/ Instrument settings menu option).
Once a change has been made, the UPDATE button must be
pressed in order to download the information to the detector.
Note:
The L.E.D intensity is set for
ALL
methods used on the detector
and is
NOT
Method specific.
3.5.3
Control Software-System Test
The System Test is used to evaluate the ELSD-1080
performance to ensure it is operating to the high standards set
by Biotage. When you select this feature (TOOLS
/ Run system
test menu option), the software will run the ELSD through a
number of internal diagnostic tests, and then compare the
results with the expected values set by Biotage. The test
information can subsequently be printed as a record.
38
Publication date: July 2010
Summary of Contents for ELSD-1080
Page 4: ...Biotage ELSD 1080 This page is intentionally left blank 4 Publication date July 2010 ...
Page 8: ...Biotage ELSD 1080 This page is intentionally left blank 8 Publication date July 2010 ...
Page 44: ...Biotage ELSD 1080 This page is intentionally left blank 44 Publication date July 2010 ...
Page 54: ...Biotage ELSD 1080 This page is intentionally left blank 54 Publication date July 2010 ...
Page 56: ...Biotage ELSD 1080 This page is intentionally left blank 56 Publication date July 2010 ...
Page 58: ...Biotage ELSD 1080 This page is intentionally left blank 58 Publication date July 2010 ...