Premier / Premier Plus

19

Operating Instructions

4.2 How to use the Premier / Premier Plus Scaler

The side of the tip-end shall be applied to the tooth so that the movement of the tip is parallel to the 

surface of the tooth. The first 2 mm of the tip are the most  efficient. Place the tip on the tooth sur-

face before activating the foot control pedal. The power is regulated with the foot control pedal from 

0 to 100% within each working mode.

Normal scaling rarely requires more than 50% power level of each working mode.  

However, hard to remove calculus might require a higher power setting. 

Normal scaling rarely requires more than 50% power level. 
However, hard to remove calculus might require a higher power 
setting. 

The tip is normally aimed towards the toothpocket.

With the correct power setting, appropriate pressure against 
the tooth (approx. 20 grams but not exceeding 50) and the tip 
operating parallel to the surface, treatment will be gentle, quiet 
and efficient.

If, during the treatment, a squeaking sound is heard (loud and 
dominating), the pressure against the tooth might be too low 
or the tip not parallel to the surface of the tooth.

To keep the tip working parallel 
to the surface of each tooth, it is 
important to follow the anatomy 
of the tooth.

 

 
 

 

How to use the scaler

The 

side

 of the tip-end shall be applied to 

the tooth so that the movement of the tip 
is parallel to the surface of the tooth. The 
first 2 mm of the tip are the most efficient! 
Place the tip on the tooth surface before 
activating the foot switch.

Ensure that contact between 
the tip and the tooth surface 
is maintained during scaling. 
Keep moving the tip slowly 
back and forth and let the 
instrument do the work. 
Use short and long strokes 
so that the whole surface of 
each tooth is scaled.

Normal scaling rarely requires more than 50% power level. 
However, hard to remove calculus might require a higher power 
setting. 

The tip is normally aimed towards the toothpocket.

With the correct power setting, appropriate pressure against 
the tooth (approx. 20 grams but not exceeding 50) and the tip 
operating parallel to the surface, treatment will be gentle, quiet 
and efficient.

If, during the treatment, a squeaking sound is heard (loud and 
dominating), the pressure against the tooth might be too low 
or the tip not parallel to the surface of the tooth.

To keep the tip working parallel 
to the surface of each tooth, it is 
important to follow the anatomy 
of the tooth.

 

 
 

 

How to use the scaler

The 

side

 of the tip-end shall be applied to 

the tooth so that the movement of the tip 
is parallel to the surface of the tooth. The 
first 2 mm of the tip are the most efficient! 
Place the tip on the tooth surface before 
activating the foot switch.

Ensure that contact between 
the tip and the tooth surface 
is maintained during scaling. 
Keep moving the tip slowly 
back and forth and let the 
instrument do the work. 
Use short and long strokes 
so that the whole surface of 
each tooth is scaled.

Ensure that contact between the tip and the tooth surface is maintained during scaling. 

Keep moving the tip slowly back and forth and let the instrument do the work. Use short 

and long strokes so that the whole surface of each tooth is scaled. The tip is normally 

aimed towards the toothpocket.

To keep the tip working parallel to the surface of each tooth, it is important to follow 

the anatomy of the tooth. With the correct power setting, appropriate pressure against 

the tooth  (approximately 20 grams but not exceeding 50) and the tip operating paral-

lel to the surface, treatment will be gentle, quiet and efficient. If, during the treatment, 

a squeaking sound is heard (loud and dominating), the pressure against the tooth might 

be too low or the tip is not parallel to the surface of the tooth.

4.3 After Treatment 

Automatic cleaning function

1. Make sure the water control is fully open. Place the scaler handpiece 

over the cuspidor bowl or sink and start the cleaning cycle by press-

ing the cleaning key “C”.

2. The cleaning cycle starts and stops automatically after 80 seconds.3. 

After the cleaning cycle is finished, screw off the tip, counter  

clockwise, with the torque wrench. 

Note

Always set the water flow to 

fully open before running the 

cleaning cycle (dots in line, 

see picture).

6

Safety notices in this manual

Warning

 indicates a potentially 

dangerous situation. Non-observ-

ance may lead to death or injury. 

Caution

 indicates a potentially 

harmful situation. Non-observ-

ance may damage the equipment. 

Note

 indicates a situation where 

special notice should be ob-

served.

WARNING 

Caution

Note

Summary of Contents for Suvi Premier

Page 1: ...Ultrasonic Piezo Scaler Place Disk Owner s Guide...

Page 2: ...53 BioSonic Suvi Premier Plus Medicament feeder AUS Plug Type I 230 60014254 Shaded and italicized items in the table above are not currently offered for sale Important Read this manual carefully befo...

Page 3: ...Premier Premier Plus 3...

Page 4: ...4 2 How to use the scaler 19 4 3 After treatment 19 Automatic cleaning function 19 5 Cleaning and Maintenance 20 5 1 Cleaning of the equipment and components 20 Cleaning 20 Disinfecting 20 Drying 20...

Page 5: ...Premier Premier Plus 5...

Page 6: ...EMC information provided in section 8 2 Portable and mobile RF Radio Frequency communications equipment can affect the product Contraindications Do not use the scaler on patients with cardiac pacemak...

Page 7: ...ning indicates a potentially dangerous situation Non observance may lead to death or injury Caution indicates a potentially harmful situation Non observance may damage the equipment Note indicates a s...

Page 8: ...features of the equipment read Equipment description section 2 6 Caution indicates a potentially harmful situation Non observ ance may damage the equipment Note indicates a situation where special not...

Page 9: ...soft silicone removable grips give the user a comfortable relaxed grip as well as an excellent feel BioSonic Suvi Premier Premier Plus is highly adaptable to any procedure or user approach It is not o...

Page 10: ...POWER OFF key 4 Working mode 1 key 0 to 40 power 5 Working mode 2 key 0 to 70 power 6 Working mode 3 key 0 to 100 power 7 StandBy indicator 8 POWER ON indicator 9 Dry mode indicator 10 Irrigation mod...

Page 11: ...12 OFF position 13 ON position 14 Diagnostic function turn LED light ON or OFF by a single click on the pedal 15 Irrigation position 16 Zero power position 17 Maximum power position 18 Power regulati...

Page 12: ...The medicament bottle can be used for either medicament solutions or ordinary clean water The unit contains an electrically driven air compressor When operating the unit the compressed air forces the...

Page 13: ...unit Caution Consult accompanying documents Type B applied part according to the degree of protection against electrical shock Fuse Input Output This appliance is labeled in accordance with European...

Page 14: ...zed Representative IPX1 Ingress Protection Rating Device has dripping water protection Date of Manufacture Autoclave at 132o C Caution Federal United States laws restricts this device to sale by or on...

Page 15: ...unit in the immediate vicinity of sources of electromagnetic radiation for example the Perfect TCS II or other electrosurgery equipment Connect the foot pedal cable to the foot control pedal and at th...

Page 16: ...t the voltage rating on the rear side matches the voltage of the AC power outlet Verify that the AC power outlet is provided with a protective ground Connect the power cord to the unit and the AC powe...

Page 17: ...ormance and must be exchanged Prolonged use may cause tip breakage and injury to the patient The operator should be aware that ultrasonic instruments with small diameters are subject to breakage at an...

Page 18: ...the ON button activates the dry mode for scaling without water medicament 7 A third press of the ON button activates the irrigation mode The irrigation mode can also be activated in the scaling mode b...

Page 19: ...t a squeaking sound is heard loud and dominating the pressure against the tooth might be too low or the tip not parallel to the surface of the tooth To keep the tip working parallel to the surface of...

Page 20: ...ar broken tip or distortion is noted 3 Clean thoroughly using an ultrasonice cleaner such as the Colt ne Whaledent BioSonic Ultrasonic Cleaning System with BioSonic UC32 Solution Concentrate Disinfect...

Page 21: ...o rinse the lines from cleaning agent solution put clean water in the bottle and run the automatic cleaning cycle until clean water comes out of the handpieces Sterilizing Package in an FDA approved w...

Page 22: ...s 1 Disconnect the power cord from the AC power outlet and the unit 2 Open the fuse holder on the rear side of the unit 3 Inspect the fuses for damage Replace damaged fuses with new ones Verify the fu...

Page 23: ...nd the fuse panel are OK 4 If the problem still remains contact your dealer for support and indicate Error Code E X02 Type of problem Please see A Unit is not responding and no lights are lit on the c...

Page 24: ...ration 1 Check that the tip is firmly tightened Preferably use a torque wrench and tighten clockwise until the torque wrench slides The torque wrench prevents the tip from being over tightened 2 Check...

Page 25: ...y another tip 6 If the problem still remains contact your dealer for support and indicate Error Code E S04 F Tip does not fit smoothly onto the handpiece 1 Clean handpiece threads with compressed air...

Page 26: ...00 g Voltage 100 Vac 50 60 Hz 115 Vac 50 60 Hz 230 Vac 50 60 Hz Primary fuse T500 mAH 250 V 5x20 mm 100 Vac T400 mAH 250 V 5x20 mm 115 Vac T200 mAH 250 V 5x20 mm 230 Vac Power cord Separate with prote...

Page 27: ...angurinara extract Sodium hypochlorite 1 solution WARNING Immedately after using any kind of medicament in the medicament bottle run the automatic cleaning cycle with clean water in the medicament bot...

Page 28: ...nt products are manufactured from new parts or new and serviceable used parts Regardless our warranty terms apply The products are guaranteed in accordance with the terms of the written Certificate of...

Page 29: ...antability and fitness for a particular purpose In no event shall C W be liable for direct indirect tort exemplary punitive special incidental or consequential damages arising from the possession or u...

Page 30: ...output lines 1 kV differential mode 2 kV common mode NOTE U T is the a c mains voltage prior to application of the test level Power frequency magnetic fields should be at levels characteristic of a ty...

Page 31: ...ic immunity Immunity test IEC 60601 test level Compliance level Electromagnetic environment guidance Portable and mobile RF communications equipment should be used no closer to any part of the BioSoni...

Page 32: ...nt in which radiated RF disturbances are controlled The customer or the user of the equipment can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobil...

Page 33: ...i Grip Color Lilac 4 60018765 Suvi Grip Color Pink 4 60018766 Suvi Grip Color Blue 4 60019328 Subgingival Nozzel 1 60018486 Polisher Nozzle Hooked 1 60018485 Polisher Nozzle Universal 1 60014456 Torqu...

Page 34: ...283 PE 40R General Scaling 60014284 PE 40H General Scaling 60019608 PE 40HS General Scaling 60014285 PE 41PS General Scaling 60014286 IM 1 Scaling Implant maintenance 4 pk 60014287 IM 2 Scaling Implan...

Page 35: ...Premier Premier Plus 35...

Page 36: ...ne Whaledent AG Feldwiesenstrasse 20 9450 Altst tten Switzerland Tel 41 0 71 757 53 00 Fax 41 0 71 757 53 01 info ch coltene com Colt ne Whaledent GmbH Co KG Raiffeisenstrasse 30 89129 Langenau Germa...

Reviews: